Advertisement
GoInductry DoveBid
GoInductry DoveBid

Integra Biosciences MEDIACLAVE

Share:

MEDIACLAVE

The MEDIACLAVE product range allows the rapid and gentle sterilisation of 1 – 30 L culture medium. Precise controlling and monitoring of temperature, time and pressure during the sterilisation process guarantee constant high quality. The intuitive graphical user interface and the simple programming make it very easy for everyone to operate the MEDIACLAVE.

Models

MEDIACLAVE 10
Attributes

This model was added by: LabWrenchEditor

Edit Model

Chamber Depth 203 mm
Orientation Benchtop
Chamber Volume 10 L
Internal Diameter 254 mm
Depth 640 mm
Height 480 mm
Width 550 mm
Power Requirements 200 – 240 V 50/60 Hz, 16 A
Weight 57 kg
Additional Specifications Capacity 
Stainless steel cuvette: 1 – 10 L  
 
Temperature range 
Sterilisation: 70°C – 122°C  
Dispensing: 30°C – 75°C  
Water bath: 30°C – 80°C  
Max. temperature deviation: +1.0°C/-0.2°C  
 
Stirrer speed 
Selectable: 50 – 200 rpm, reversing direction  
 
Utilities 
Heating Capacity: Max. 3 kW  
Interfaces: 2 x RS232, Ethernet, AUX contact, footswitch, ext. pinch valve  
Cooling water connections: ¾" thread
MEDIACLAVE 30
Attributes

This model was added by: LabWrenchEditor

Edit Model

Orientation Floor Standing
Chamber Volume 30 L
Depth 640 mm
Height 1040 mm
Width 550 mm
Power Requirements 200-208 V 3~/PE, 50/60 Hz, 30 A
Weight 85 kg
Additional Specifications Capacity 
Stainless steel cuvette: 3 – 30 L 
 
Temperature range 
Sterilisation: 70°C – 122°C  
Dispensing: 30°C – 75°C  
Water bath: 30°C – 80°C  
Max. temperature deviation: +1.0°C/-0.2°C  
 
Stirrer speed 
Selectable: 50 – 200 rpm, reversing direction  
 
Utilities 
Heating Capacity: Max. 9 kW  
Interfaces: 2 x RS232, Ethernet, AUX contact, footswitch, ext. pinch valve  
Cooling water connections: ¾" thread

Features

Operation modes:        
Two main operation modes for media preparation, plus two extra modes can be selected and the parameters set to your needs:        
        
1. STANDARD mode: For the preparation of standard an highly sensitive culture media. Sterilisation temperature/time and dispensing temperature can be set.        
        
2. CHOCOLATE AGAR mode: A special two-step program that allows the preparation of complex media. After the first sterilisation phase, supplement, e.g. blood, can be added through the adding port. Subsequently, a second heating phase will be performed.        
        
3. WATER BATH mode: For pre-swelling culture media prior to sterilisation within the temperature range 30 – 80°C. In combination with an autoclave cuvette (only MEDIACLAVE 10), liquids in glassware can be thermostatted.         
        
4. AUTOCLAVE mode (MEDIACLAVE 10 only):        
MEDIACLAVE 10, together with an optional autoclave kit can be used as a bench top autoclave for the sterilisation of small amounts of culture media in glassware such as Erlenmeyer flasks or test tubes.

Photos

MEDIACLAVE

MEDIACLAVE

Manuals and Resources

Note: We appreciate the support of individuals and companies making their documents and resources available to our community. If you find a document or resource that does not belong please let us know by emailing info@labwrench.com

Instrument Documents

Picture of Integra Biosciences MEDIACLAVE
INTEGRA is a leading provider of high-quality laboratory tools for liquid handling, media preparation, sterilisation and cell cultivation. We are committed...

More about Integra Biosciences

Visit Integra Biosciences

Questions and Answers

Recent Question Activity

Ask a Question

The Mediaclave is blocked

after we have traced and checked it's overtemp protector and found out the overtemp was still ok, the message still remains. how do we unblock or reset the equipment?

There are 1 Threads

Reviews

Member Reviews

There are no threads in Integra Biosciences MEDIACLAVE Reviews at this time.
Add a Review