LABSYS evo

LABSYS evo offers TGA with simultaneous DTA or DSC and Cp measurement within the temperature range of ambient to 1 600 °C. The LABSYS evo features a NEW high performance top loading balance and a unique 3D high-precision (Calvet type) Cp rod. Thanks to its unique ergonomic design and market leading Calisto software the LABSYS evo offers you the most easy-to-use Thermal Analysis system with uncompromised performance.
LABSYS evo
Attributes
Cooling Rate .01 to 100 °C/min
Heating Rate .01 to 100 °C/min
Temperature Range 28 to 1600 °C
DSC Type Heat Flux DSC
Depth 22.8 in
Height 22 in (closed)
Width 20.8 in
Power Requirements 230 V - 50/60 Hz
Weight 121 lbs
Additional Specifications Temperature Range:     
Ambient to 1 600 °C    
Isothermal temperature accuracy:     
+/- 1 °C    
Isothermal temperature precision:     
+/- 0.04 °C    
Programmable temperature scanning rate (heating and cooling:     
0.01 to 100 °C.min-1    
Furnace cooling:     
32 min (1600°C to 50°C)    
Maximum balance capacity:     
20 g    
Weight range:     
+/- 1 000 mg - +/- 200 mg    
Weighing precision:     
+/- 0.01 %    
TG resolution:     
0.2 Μg - 0.02 Μg    
Baseline dynamic drift (one hour):     
10 Μg    
TG reproductibility up to 1 200 °C:     
< 50 Μg    
DSC rod - Resolution:     
0.4 µW / 10 ΜW    
3D Cp rod - Cp Accuracy:     
< 2 %    
Vacuum:     
< 10-1 mbar    
Evolved gas:     
simultaneous MS, FTIR couplings (option)    
Gases:     
Two inlets for gas scanning (inert or reactive) OR 3 carrier gases (option) (MFC from 4 to 200 ml/min) + 1 auxiliary or reactive gas (MFC from 0.3 to 16 ml/min)

Features

Simplicity of use: thanks to both an ergonomic design and the dedicated market leading Calisto software, LABSYS evo is user friendly and intuitive to operate    
    
Wide temperature range: TGA with simultaneous DTA or DSC and Cp measurement can be achieved for temperature up to 1 600 °C    
    
High Precision Cp Measurement: the unique 3D (Calvet type) multiple thermocouples design enables large mass to be measured and to perform a Cp within 2 % , throughout the whole temperature range (ambient – 1 600 °C)    
    
New top loading balance: the unmatched stability, reproducibility and accuracy of this new balance offer unparalleled TGA data

Photos

LABSYS evo

LABSYS evo

Manuals and Resources

Note: We appreciate the support of individuals and companies making their documents and resources available to our community. If you find a document or resource that does not belong please let us know by emailing info@labwrench.com

Instrument Documents

Picture of Setaram Instrumentation LABSYS evo
For fifty years, Setaram Instrumentation (Caluire, France) has been the global supplier of solutions in thermal analysis and calorimetry, a domain of...

More about Setaram Instrumentation

Visit Setaram Instrumentation

Questions and Answers

Recent Question Activity

Ask a Question

LABSYS evo

We have LAbsys Evo TGA in our lab. I'm facing problem in the results. After start the TGA the weight should be decreased but in my case, there is a slight increase in the weight than decreased as normal. I don't know how to fix it?

There are 1 Threads

Reviews

Member Reviews

There are no threads in Setaram Instrumentation LABSYS evo Reviews at this time.
Add a Review

Buy it Used!

Used Setaram Instrumentation LABSYS evo Offerings on LabX

LABTECH MV5 Automated Nitrogen Concentrator

Price: Please Email

Condition:

Configuration: As Shown

Used Labsys EVO Removed from Local Lab in Working Condition

Price: $45,000.00

Condition:

Configuration: As Shown

Advertisement

Videos

See it in action

Userfriendly Thermal Analysis Platform LABSYS evo with Auto Sampler

LABSYS evo is the most userfriendy thermal analysis platform that offers TGA, STA (Simultaneous Thermal Analysis), DTA and DSC measurements up to...