Hach Company DR 6000

Share:

DR 6000

The DR6000™ is the industry’s most advanced lab spectrophotometer. It offers high speed wavelength scanning across the UV and Visible Spectrum, and comes with over 250 pre-programmed methods which include the most common testing methods used today. With optional accessories allowing for high volume testing via a carousel sample changer, and increased accuracy with a sample delivery system which eliminates optical difference errors, this instrument ensures you are ready to handle your wide-ranging water testing needs. 
 
Combining the DR6000 with Hach’s patented TNTplus™ reagent vials, you gain additional accuracy with guided step-by-step testing procedures, while making scratched, flawed or dirty glassware a non-issue due to 10 reading averages and the discarding of outliers. The instrument also utilizes RFID technology, allowing it to automatically let you know if your TNTplus reagents have expired, while detecting coefficient factors to avoid errors which can occur in lot-to-lot variations in the chemistries. 
 
Partnering all of these features with an integrated QA software package allows for scheduling, documenting and interpreting your needed quality measurements. 
 
Your water testing needs, all in one spectrophotometer. 
 
*RFID currently available only in US, Canada, Puerto Rico, Australia, New Zealand, and Colombia. Customers in other countries should order LPV441.99.00002.

Models

DR 6000 with RFID Technology
Attributes

Edit Model

Bandwidth 2 nm
Light Source Tungsten (visible range), deuterium (UV range)
Photometric Range ± 3 Abs
Scanning Speed 900 nm/min (in 1 nm steps
Stray Light KI-solution at 220 nm < 3.3 Abs/ < 0.05%
Wavelength Range 190 to 1100 nm
Wavelength Resolution .1 nm
Depth 460 mm
Height 215 mm
Width 500 mm
Power Requirements 100 - 240 V; 50/60 Hz
Weight 11 kg
Part Number LPV441.99.00012
Additional Specifications Operating Conditions: 10 to 40 °C, max. 80% relative humidity (non-condensing)
Wavelength Accuracy: ± 1 nm
Wavelength Reproducibility: < 0.1 nm
Wavelength Selection: Automatic, based on method selection
Photometric Accuracy: <1% at 0.5-2.0 Abs at 546 nm 5 mAbs at 0.0-0.5 Abs
Photometric Linearity: 0.5 % - 2 Abs
Photometric Linearity 2: 1 % at > 2 Abs with neutral glass at 546 nm
Optical System: reference beam, spectral
Manual Languages: en, es, fr, ja, ko, pt, zh
DR6000 without RFID Technology
Attributes

Edit Model

Light Source Tungsten (visible range), deuterium (UV range)
Stray Light KI-solution at 220 nm < 3.3 Abs/ < 0.05%
Wavelength Range 190 to 1100 nm
Wavelength Resolution .1 nm
Depth 460 mm
Height 215 mm
Width 500 mm
Power Requirements 100 - 240 V
Weight 11 kg
Part Number LPV441.99.00002
Additional Specifications Cuvette Compatibility 2: 1 inch rectangular and round cell
Cuvette Compatibility 3: optional 100 mm rectangular cell with additional adapter
Operating Conditions: 10 to 40 °C, max. 80% relative humidity (non-condensing)
Wavelength Accuracy: ± 1 nm
Wavelength Reproducibility: < 0.1 nm
Wavelength Selection: Automatic
Photometric Accuracy: 5 mAbs at 0.0-0.5 Abs
Photometric Linearity: 0.5 % - 2 Abs
Photometric Linearity 2: 1 % at > 2 Abs with neutral glass at 546 nm
Optical System: reference beam, spectral
Manual Languages: en, es, fr, ja, ko, pt, zh

Features

Your Water Testing Needs, All in One Spectrophotometer 
The DR 6000 has the most pre-programmed testing methods, including high speed wavelength scanning across the UV and Visible Spectrum. 
 
Accessories for High Volume and High Accuracy Testing Needs 
A carousel sample changer allows for up to seven sequential measurements. The Sipper Module, an instrument controlled sample delivery system, increases precision by constant optical characteristics. 
 
Advanced Quality Assurance at Your Fingertips 
The DR 6000 comes with integrated QA software for scheduling, documenting and interpreting all of your needed quality measurements. 
 
Guided Procedures and Elimination of False Readings 
The DR 6000, when used with TNTplus® reagent vials provides you the accurate results you need by guiding you step-by-step through your testing procedures. With TNTplus the instrument averages 10 readings and eliminating outliers, making scratched, flawed or dirty glassware a non-issue. 
 
Automatically Avoids Errors 
RFID* technology automatically updates the program calibration factors when you place a TNTplus reagent box near the DR 6000. The instrument identifies chemistry expiration dates via a barcode on the vials, and detcts chemistry coefficient factors to avoid errors that can occur in lot-to-lot variations in the chemistry. 
*RFID technology currently available in US, Canada, Australia and New Zealand only.

Photos

DR 6000 UV VIS Spectrophotometer

DR 6000 UV VIS Spectrophotometer

Manuals and Resources

Note: We appreciate the support of individuals and companies making their documents and resources available to our community. If you find a document or resource that does not belong please let us know by emailing info@labwrench.com

Picture of Hach Company DR 6000
HACH COMPANY manufactures and distributes analytical instruments and reagents used to test the quality of water and other aqueous solutions. Our systems...

More about Hach Company

Visit Hach Company

Questions and Answers

Recent Question Activity

Ask a Question

DR6000 working principle

My company using DR6000 but I wonder why we need to do a standard check for most of the test methods. Is it necessary? With the initialisation in this instrument, it should have done checks on detector n the rest right?

There are 1 Threads

Reviews

Member Reviews

There are no threads in Hach Company DR 6000 Reviews at this time.
Add a Review

Hach Company DR 6000 Articles

Helpful Resources, Articles, and Information

Hach Company has developed the latest innovation in lab testing, the DR 6000™ UV-Vis spectrophotometer. The DR 6000 is equipped with RFID...

See All Related Articles