Hewlett Packard 5890 Series II

5890 Series II

The HP 5890 Series II provides flexibility in choices among inlets, columns, and detectors through the use of liners and adapters, allowing any standard column to be used without sacrificing performance. Additional flexibility is gained through positions of inlets and detectors relative to each other and through the large internal volume of the oven. There are five inlet options and six detector, as well as a wide variety of flow and pressure control components.

Instrument Attributes
Additional Specifications Heated Devices
  • Five heated zones standard: Two detectors, Two inlets, one auxiliary
  • Methods stored: Two

Split/Splitless Capillary
  • Range to 400°C in 1°C increments
  • Back-pressure design permits independent adjustment of split flow rate without
    affecting column flow
  • Septum purge built in at 3 ml/min
  • Accepts columns up to 1.2-mm od
  • Multimode design includes split and splitless injection
  • Splitless purge time variable in 0.01-min elements

Flame Ionization Detector
  • Range to 450°C
  • Grounded jet and current limited design for operator safety
  • Push-button flame ignition
  • Minimum detectable: 5 pg carbon/sec, nitrogen carrier at S/N=2
  • Linear dynamic range: +/- 10% over a 10 7 range with 0.018-in. id jet
  • Conditions: column flow 50 ml/min, 45 ml/min H2, 650 ml/min air, propane
    sample

Thermal Conductivity Detector
  • Range to 400 °C
  • Single-filament (single column) design has fluidic switching of reference and
    analytical carrier flows
  • Minimum detectable :400 pg/ml carrier -- equivalent to 1 pmm neon in 1 ml air
  • Linear dynamic range:+/- 5% over 105 range
  • Digital gain setting time-programmable through the keyboard Conditions:
    detector
  • 100 °C, 45 ml/min switching and 30 ml/min analytical flow of helium, propane
    sample

Features

Common Configurations:
  • 5890 Series II single injector (split/split less or Packed) single FID detector
  • 5890 Series II single injector (split/split less or Packed) single TCD detector
  • 5890 Series II dual injectors (split/split less or Packed) dual FID detectors
  • 5890 Series II dual injectors (split/split less or Packed) dual TCD detectors
  • 5890 Series II dual injectors (split/split less or Packed) dual FID/TCD detectors
  • All of the above systems come with INET communications (HPIB and data system is
    available.)

Photos

  • PHOTO_1
PHOTO_1

PHOTO_1

Manuals and Resources

Instrument Documents

Link to more information

Picture of Hewlett Packard 5890 Series II

Hewlett Packard

Hewlett-Packard became Agilent Technologies in November 1999.

Agilent Life Sciences and Chemical Analysis is a leading global provider of instrumentation,...

More about Hewlett Packard

This instrument was added by: admin

last edited by: WrenchAdmin

Questions and Answers

Recent Question Activity

Ask a Question

reading not stable

our gc use capillary to detect low concentration of hydrogen sulfide. reading for sample not related with standard. reading standard also difficult to get stable data.

sample size

If you try to manually inject 0.2 microliters of the same chemical 3 or 4 times in a row, and your total area varies from ~220,000,000 to 120,000,000 - does this mean your sample size is larger one time than the next? What can I do?

unidentified peaks

I'm new to this so bare with me, I have 2 HP5890's and I'm having problems with peak identifications. I usually work with GC/MS with a NIST Library, but these units are FID. I understand about running standards to define retention times, but what about samples

Preventive Maintenance

Keep your instrument running smooth

There are no preventive maintenance tips for Hewlett Packard 5890 Series II

General Comments

Post your thoughts

Application Note: The Analysis of Chlorinated Pesticides and PCBs Using The HP-608 Capillary Column

AN 228-236 : The Analysis of Chlorinated Pesticides and PCBs Using The HP-608 Capillary Column

Abstract: This 9-page application note compares splitless and on- column injection techniques using the HP 5890 Series II GC with EPC interface board.

Popular low cost, single instrument chromatography data system range

News 02/2004 - Chromatography Data System Software for Agilent 5890 and 6890 Gas Chromatographs includes full instrument/autosampler control.
Justice Laboratory Software, the leading developer of chromatography data system software for science based

Buy it Used!

Used Hewlett Packard 5890 Series II Offerings on LabX

HEWLETT-PACKARD 5890 SERIES II GC WITH 7673B

Price: 9,500

Condition: Excellent

Configuration: Complete

HP Gas Chromatagraph 5890 Series II FID/TCD

Price: best offer

Condition: Fair/Working

Configuration: As Shown

HEWLETT-PACKARD 5890 SERIES II GC WITH 7673B

Price: 9,500

Condition: Excellent

Configuration: Complete

Lab Wrench Community