Advertisement

Beckman Coulter Avanti J Series

Share:

Avanti J Series

The Beckman Coulter Avanti® J Series features performance capabilities unattainable with high speed or super speed centrifuges. New rotors enable forces of over 110,000 g. Features include a high-torque SR Drive* for the fastest acceleration and deceleration possible with gentle precision. All models feature automatic rotor identification and imbalance safety check, CFC-free refrigerants, FRS (Friction Reduction System), and accommodate all Beckman Coulter J rotors.          
          
The Avanti J-30I High-Performance Centrifuge System gives you the fastest separations possible - in the shortest amount of time. Swinging-bucket and fixed-angle rotors provide maximum separation forces in excess of 100,000 x g at speeds up to 30,000 rpm. Your productivity is ensured by unmatched accel/decel rates and an easy run set-up. Your sample spends more time at full force and less time in the centrifuge.

Models

Avanti J-30I
Attributes

This model was added by: WrenchHelp

Edit Model

Centrifuge Type High Speed
Maximum Capacity 4 liters (Single tube Volume Range 1.5 - 1,000 mL)
Maximum RCF 110500 xg
Maximum Speed 30000 rpm
Noise Level < 57 dB
Orientation Floor Standing
Refrigerated Yes_ non-ozone depleting refrigerant
Speed Range 0 to 30000 rpm 0 to 6,000 rpm for elutriation
Depth 34.5 in
Height 34.1 in
Width 28 in
Weight 680 lbs
Additional Specifications Approximate Acceleration Time: Maximum, slow, or timed (1 to 10 min from 0 to 500 rpm)   
Approximate Deceleration Time: Maximum, slow, timed (1 to 10 min from 500 to 0 rpm), or off   
Friction Reduction System (FRS): A mechanical rotary vane pump reduces chamber pressure dependent on rotor and run speed   
Maximum Heat Dissipation into Room under Steady-State Conditions: 2.0 kW (6,900 Btu/hr)   
Set Speed: Actual rotor speed ± 10 rpm from 10,000 to 30,000 rpm or equivalent RCF   
Set Time: 1 min to 99 hr 59 min or continuous (hold)   
Temperature Control: ± 2°C of set temperature (after equilibration)   
Ambient Temperature Range: 16 to 38°C   
Clearances Required: 7.6 cm (3.0 in) sides; 16.0 cm (6.25 in) rear   
Door: 5.4-cm (2.1-in) thick structural foam with steel plate   
Humidity Restrictions: < 95% (noncondensing)   
Pollution Degree: 2   
Rotor Chamber Diameter: 40.6 cm (16 in)   
Time -- Actual Display: Indicates run time remaining (timed run) or elapsed time (hold run)
Avanti J-26 XP
Attributes

This model was added by: WrenchHelp

Edit Model

Centrifuge Type High Speed
Maximum Capacity 6 liters (single tube volume range: 1.5 - 1,000 mL)
Maximum RCF 82000 xg
Maximum Speed 26000 rpm
Noise Level < 57 dB
Orientation Floor Standing
Refrigerated No_ non-ozone depleting refrigerant
Speed Range 100 to 26000 rpm or equivalent RCF
Depth 34.5 in
Height 34.1 in
Width 28 in
Weight 640 lbs
Additional Specifications Approximate Acceleration Time: Maximum, slow, or timed (1 to 10 min from 0 to 500 rpm)   
Approximate Deceleration Time: Maximum, slow, timed (1 to 10 min from 500 to 0 rpm), or off   
Drive Cooling: Air Cooled   
Drive Type: SR* Drive   
Friction Reduction System (FRS): A mechanical rotary vane pump reduces chamber pressure to 0.25 ATM (190 mm [7.5 in] Hg)   
Maximum Heat Dissipation into Room under Steady-State Conditions: 2.0 kW (6,900 Btu/hr)   
Rotor ID: Auto   
Set Speed: Actual rotor speed ± 10 rpm from 10,000 to 26,000 rpm or equivalent RCF   
Set Time: 1 min to 99 hr 59 min, 2t, or continuous (hold)   
Speed Control: ± rpm of set speed or 0.1%, whichever is greater   
Speed Range: 0 to 5,000 rpm for elutriation   
Temperature Control: ± 2°C of set temperature (after equilibration)   
Zonal Capability: Yes   
Ambient Temperature Range: 15 to 40°C
J-26 XPI
Attributes

This model was added by: WrenchHelp

Edit Model

Centrifuge Type High Speed
Maximum Capacity 6 liters (single tube volume range: 1.5 - 1,000 mL)
Maximum RCF 82000 xg
Maximum Speed 26000 rpm
Noise Level < 57 dB
Orientation Floor Standing
Refrigerated No_ non-ozone depleting refrigerant
Depth 34.5 in
Height 34.1 in
Width 28 in
Weight 340 lbs
Additional Specifications Accel/Decel Rates 12/13  
Temp Control + 2° of set temperature  
Speed Control + 10 rpm of set speed or 0.1%, whichever is greater  
Ambient Temp Range 16° to 38°C  
Drive Type Cooling SRDrive / Air Cooled  
Power Usage / Heat Output 2.0 kW / 6900 Btu/hr
Avanti J-E
Attributes

This model was added by: WrenchHelp

Edit Model

Centrifuge Type Ultraspeed
Maximum Capacity 4 liters (single tube volume range: 1.5 - 1,000 mL)
Maximum RCF 53000 xg
Maximum Speed 21000 rpm
Noise Level < 64 dB
Orientation Floor Standing
Refrigerated Yes CFC-Free non-ozone depleting refrigerant
Speed Range 300 to 21000 rpm
Depth 31.5 in
Height 36 in
Width 25 in
Power Requirements 50 Hz, 180 to 253 VAC, 24 A; 60 Hz, 187to229/216to264 VAC, 24 A
Weight 589 lbs
Additional Specifications Approximate Acceleration Time: Two profiles--MAX (full acceleration to set speed) and SLOW
Approximate Deceleration Time: Three profiles--MAX, SLOW, and OFF (coast)
Friction Reduction System (FRS): 190 mm (7.5 in) Hg
Maximum Heat Dissipation into Room under Steady-State Conditions: 2.0 kW (6,900 Btu/hr)
Set Speed: Actual rotor speed ± 50 rpm of set speed
Set Time: to 99 hr 59 min or continuous (hold)
Temperature Control: ± 2° C of set temperature (after equilibration)
Ambient Temperature Range: 15° C to 35° C
Clearances Required: 7.6 cm (3.0 in) sides; 16.0 cm (6.25 in) rear
Humidity Restrictions: < 95% (noncondensing)
Pollution Degree: 2
Time -- Actual Display: Indicates run time remaining/displays elapsed time

Finish
Control panel: silicone elastomer.
Top surface: urethane paint.
Other surfaces: general purpose paint.

Features

Near-Ultracentrifuge Forces          
              
  • Over 100,000 x g means fast protein separations in the Avanti J-30I and the JA-30.50, the highest g-force fixed-angle rotor available in a high-performance centrifuge          
  • Subcellular organelles effectively separated by rate zonal centrifugation with the JS-24.38, a 6 x 38 mL swinging-bucket rotor          
          
High Throughput          
              
  • Four-liter batch throughput for bacteria and cell membrane isolation using the JLA-9.1000 J-LITE® rotor at 16,800 x g          
  • DNA sample preparation in up to 10 microplates with the JS-5.9 Rotor           
          
Fast and Efficient          
              
  • Fastest accel and decel times with exclusive SR Drive** technology          
  • Low heat output and low energy consumption          

Photos

  • Avanti J-HC
  • J6-MI

Manuals and Resources

Note: We appreciate the support of individuals and companies making their documents and resources available to our community. If you find a document or resource that does not belong please let us know by emailing info@labwrench.com

Instrument Documents

Picture of Beckman Coulter Avanti J Series
Beckman Coulter, Inc. is a leading manufacturer of biomedical testing instrument systems, tests and supplies that simplify and automate laboratory processes....

More about Beckman Coulter

Visit Beckman Coulter

Questions and Answers

Recent Question Activity

Ask a Question

J-30I D4 error

Does anyone know how to fix a D4 error on an Avant J-30I centrifuge? Please help! Thanks, Paul 

Sys error Avanti J-26 XP

Hello, We have a Beckman Coulter Avanti J-26 XP in our lab. Last Friday it gave the error "sys" when we switched it on and it was no possible to delete it by pressing "clear". This morning I have tried a sequence of keys I...

Beckman AvantiTM J-25 Centrifuge with "Door" LED flashing

I have a Beckman AvantiTM J-25 Centrifuge. It has "Door" LED flashing when power is switched on. This happened after it was relocated from one building to another building. I tried pressing "CLEAR"  then "FRS" also...

See all 24 Beckman Coulter Avanti J Series Questions and Answers There are 24 Threads

Reviews

Member Reviews

There are no threads in Beckman Coulter Avanti J Series Reviews at this time.
Add a Review