Beckman Coulter Avanti J Series

Share:

Avanti J Series

The Beckman Coulter Avanti® J Series features performance capabilities unattainable with high speed or super speed centrifuges. New rotors enable forces of over 110,000 g. Features include a high-torque SR Drive* for the fastest acceleration and deceleration possible with gentle precision. All models feature automatic rotor identification and imbalance safety check, CFC-free refrigerants, FRS (Friction Reduction System), and accommodate all Beckman Coulter J rotors.          
          
The Avanti J-30I High-Performance Centrifuge System gives you the fastest separations possible - in the shortest amount of time. Swinging-bucket and fixed-angle rotors provide maximum separation forces in excess of 100,000 x g at speeds up to 30,000 rpm. Your productivity is ensured by unmatched accel/decel rates and an easy run set-up. Your sample spends more time at full force and less time in the centrifuge.

Models

Avanti J-30I
Attributes

This model was added by: WrenchHelp

Edit Model

Centrifuge Type High Speed
Maximum Capacity 4 liters (Single tube Volume Range 1.5 - 1,000 mL)
Maximum RCF 110500 xg
Maximum Speed 30000 rpm
Noise Level < 57 dB
Orientation Floor Standing
Refrigerated Yes_ non-ozone depleting refrigerant
Speed Range 0 to 30000 rpm 0 to 6,000 rpm for elutriation
Depth 34.5 in
Height 34.1 in
Width 28 in
Weight 680 lbs
Additional Specifications Approximate Acceleration Time: Maximum, slow, or timed (1 to 10 min from 0 to 500 rpm)   
Approximate Deceleration Time: Maximum, slow, timed (1 to 10 min from 500 to 0 rpm), or off   
Friction Reduction System (FRS): A mechanical rotary vane pump reduces chamber pressure dependent on rotor and run speed   
Maximum Heat Dissipation into Room under Steady-State Conditions: 2.0 kW (6,900 Btu/hr)   
Set Speed: Actual rotor speed ± 10 rpm from 10,000 to 30,000 rpm or equivalent RCF   
Set Time: 1 min to 99 hr 59 min or continuous (hold)   
Temperature Control: ± 2°C of set temperature (after equilibration)   
Ambient Temperature Range: 16 to 38°C   
Clearances Required: 7.6 cm (3.0 in) sides; 16.0 cm (6.25 in) rear   
Door: 5.4-cm (2.1-in) thick structural foam with steel plate   
Humidity Restrictions: < 95% (noncondensing)   
Pollution Degree: 2   
Rotor Chamber Diameter: 40.6 cm (16 in)   
Time -- Actual Display: Indicates run time remaining (timed run) or elapsed time (hold run)
Avanti J-26 XP
Attributes

This model was added by: WrenchHelp

Edit Model

Centrifuge Type High Speed
Maximum Capacity 6 liters (single tube volume range: 1.5 - 1,000 mL)
Maximum RCF 82000 xg
Maximum Speed 26000 rpm
Noise Level < 57 dB
Orientation Floor Standing
Refrigerated No_ non-ozone depleting refrigerant
Speed Range 100 to 26000 rpm or equivalent RCF
Depth 34.5 in
Height 34.1 in
Width 28 in
Weight 640 lbs
Additional Specifications Approximate Acceleration Time: Maximum, slow, or timed (1 to 10 min from 0 to 500 rpm)   
Approximate Deceleration Time: Maximum, slow, timed (1 to 10 min from 500 to 0 rpm), or off   
Drive Cooling: Air Cooled   
Drive Type: SR* Drive   
Friction Reduction System (FRS): A mechanical rotary vane pump reduces chamber pressure to 0.25 ATM (190 mm [7.5 in] Hg)   
Maximum Heat Dissipation into Room under Steady-State Conditions: 2.0 kW (6,900 Btu/hr)   
Rotor ID: Auto   
Set Speed: Actual rotor speed ± 10 rpm from 10,000 to 26,000 rpm or equivalent RCF   
Set Time: 1 min to 99 hr 59 min, 2t, or continuous (hold)   
Speed Control: ± rpm of set speed or 0.1%, whichever is greater   
Speed Range: 0 to 5,000 rpm for elutriation   
Temperature Control: ± 2°C of set temperature (after equilibration)   
Zonal Capability: Yes   
Ambient Temperature Range: 15 to 40°C
J-26 XPI
Attributes

This model was added by: WrenchHelp

Edit Model

Centrifuge Type High Speed
Maximum Capacity 6 liters (single tube volume range: 1.5 - 1,000 mL)
Maximum RCF 82000 xg
Maximum Speed 26000 rpm
Noise Level < 57 dB
Orientation Floor Standing
Refrigerated No_ non-ozone depleting refrigerant
Depth 34.5 in
Height 34.1 in
Width 28 in
Weight 340 lbs
Additional Specifications Accel/Decel Rates 12/13  
Temp Control + 2° of set temperature  
Speed Control + 10 rpm of set speed or 0.1%, whichever is greater  
Ambient Temp Range 16° to 38°C  
Drive Type Cooling SRDrive / Air Cooled  
Power Usage / Heat Output 2.0 kW / 6900 Btu/hr
Avanti J-E
Attributes

This model was added by: WrenchHelp

Edit Model

Centrifuge Type Ultraspeed
Maximum Capacity 4 liters (single tube volume range: 1.5 - 1,000 mL)
Maximum RCF 53000 xg
Maximum Speed 21000 rpm
Noise Level < 64 dB
Orientation Floor Standing
Refrigerated Yes CFC-Free non-ozone depleting refrigerant
Speed Range 300 to 21000 rpm
Depth 31.5 in
Height 36 in
Width 25 in
Power Requirements 50 Hz, 180 to 253 VAC, 24 A; 60 Hz, 187to229/216to264 VAC, 24 A
Weight 589 lbs
Additional Specifications Approximate Acceleration Time: Two profiles--MAX (full acceleration to set speed) and SLOW
Approximate Deceleration Time: Three profiles--MAX, SLOW, and OFF (coast)
Friction Reduction System (FRS): 190 mm (7.5 in) Hg
Maximum Heat Dissipation into Room under Steady-State Conditions: 2.0 kW (6,900 Btu/hr)
Set Speed: Actual rotor speed ± 50 rpm of set speed
Set Time: to 99 hr 59 min or continuous (hold)
Temperature Control: ± 2° C of set temperature (after equilibration)
Ambient Temperature Range: 15° C to 35° C
Clearances Required: 7.6 cm (3.0 in) sides; 16.0 cm (6.25 in) rear
Humidity Restrictions: < 95% (noncondensing)
Pollution Degree: 2
Time -- Actual Display: Indicates run time remaining/displays elapsed time

Finish
Control panel: silicone elastomer.
Top surface: urethane paint.
Other surfaces: general purpose paint.

Features

Near-Ultracentrifuge Forces          
              
  • Over 100,000 x g means fast protein separations in the Avanti J-30I and the JA-30.50, the highest g-force fixed-angle rotor available in a high-performance centrifuge          
  • Subcellular organelles effectively separated by rate zonal centrifugation with the JS-24.38, a 6 x 38 mL swinging-bucket rotor          
          
High Throughput          
              
  • Four-liter batch throughput for bacteria and cell membrane isolation using the JLA-9.1000 J-LITE® rotor at 16,800 x g          
  • DNA sample preparation in up to 10 microplates with the JS-5.9 Rotor           
          
Fast and Efficient          
              
  • Fastest accel and decel times with exclusive SR Drive** technology          
  • Low heat output and low energy consumption          

Photos

  • Avanti J-HC
  • J6-MI

Manuals and Resources

Note: We appreciate the support of individuals and companies making their documents and resources available to our community. If you find a document or resource that does not belong please let us know by emailing info@labwrench.com

Instrument Documents

Picture of Beckman Coulter Avanti J Series
Beckman Coulter, Inc. is a leading manufacturer of biomedical testing instrument systems, tests and supplies that simplify and automate laboratory processes....

More about Beckman Coulter

Visit Beckman Coulter

Questions and Answers

Recent Question Activity

Ask a Question

Beckman Avanti J30I Centrifuge (L3, L4, D4 errors)

We have J30I  (Ser. JKY05G08) and few months ago, all of sudden, it was giving me L3, L4 and D4 error.  Oddly, when I cleared the error, turned it off and disconnected the power and it was working fine.  And, recently, it came back...

avanti 30i

 hello all   i have error code D4 in AVANTI 30I centrifuge  please i need your help   thanks      rafat  

R1

See all 29 Beckman Coulter Avanti J Series Questions and Answers There are 29 Threads

Reviews

Member Reviews

There are no threads in Beckman Coulter Avanti J Series Reviews at this time.
Add a Review