When cost is a primary consideration of selection, the Series 1000 provides low cost, high specification, quality instruments for both the UV and visible ranges. The designs offer an ideal combination of performance, quality and reliability for educational use; as is witnessed by the large number of Universities, Polytechnics and Colleges who are multiple users of the range.

Models

CE 1011
Attributes

This model was added by: WrenchAdmin2

Edit Model

Bandwidth 8 nm
Optics Single Beam
Photometric Noise Less than + 0.001A/Hr.
Photometric Range -0.3 to 3.0A, 0-200%T, 0-9999C
Sample Compartment Dimensions: 196 x 105 x 107 mm (W x D x H)
Stray Light Less than 0.05% at 220nm and 340nm CE1021
Wavelength Range 325 to 1100 nm
Depth 340 mm
Height 200 mm
Width 480 mm
Power Requirements 110-250V, 50/60Hz, 120W
Weight 10 kg
Additional Specifications Wavelength Selection: Keyboard controlled by up-down keys
Wavelength Accuracy: Within 1nm
Wavelength Precision: 0.1nm
Wavelength Display: Large seven segment bright LED’s
Photometric Accuracy: 1% or 0.005A whichever is greater
Absorbance Zero Stability: Better than + 0.001A/Hr.
Absorbance Zero Setting: 0A and 100%T automatic by press button
Peak Seek: Automatic with baseline flattening
Filter and Lamp Selection: Automatic - coupled to wavelength
Self Test and Calibration: Automatic at switch on
Computer Interface: Bi-directional serial RS232c
CE 1021
Attributes

This model was added by: WrenchAdmin2

Edit Model

Bandwidth 8 nm
Optics Single Beam
Photometric Noise Less than + 0.001A/Hr.
Photometric Range -0.3 to 3.0A, 0-200%T, 0-9999C
Sample Compartment Dimensions: 196 x 105 x 107 mm (W x D x H)
Wavelength Range 200 to 1100 nm
Depth 340 mm
Height 200 mm
Width 480 mm
Power Requirements 110-250V, 50/60Hz, 170W
Weight 17 kg
Additional Specifications Wavelength Selection: Keyboard controlled by up-down keys
Wavelength Accuracy: Within 1nm
Wavelength Precision: 0.1nm
Wavelength Display: Large seven segment bright LED’s
Photometric Accuracy: 1% or 0.005A whichever is greater
Absorbance Zero Stability: Better than + 0.001A/Hr.
Absorbance Zero Setting: 0A and 100%T automatic by press button
Peak Seek: Automatic with baseline flattening
Filter and Lamp Selection: Automatic - coupled to wavelength
Self Test and Calibration: Automatic at switch on
Computer Interface: Bi-directional serial RS232c

Features

Ease of Operation

With only six, carefully programmed operating keys, the Series 1000 spectrophotometers are exceptionally easy to use. Up-down keys, with automatic slew rate acceleration, make wavelength selection, to 0.1nm, convenient and rapid. Absorbance, Transmittance or Concentration may be measured; the latter using standards or by entering a factor.

Wavelength Precision

Very precise wavelength selection, with readout to 0.1nm, provide the high degree of reproducibility needed for accuracy, particularly when measuring on a sloping background.

Convenient Operation

Design advantages of the series 1000 instruments include self test and calibration at switch on, automatic lamp and filter selection coupled to wavelength and high overall stability eliminating frequent resetting of the absorbance zero when making extended measurements.

Deuterium Lamp Save

UV-visible instruments allow the Deuterium lamp to be switched off when working in the visible region, to extend lamp life. Switch on is automatic if a UV wavelength is selected. Lamps are guaranteed for 1500 hours and are fitted with an hours of use indicator.

User Benefits
  • Easy Use
  • Low Cost
  • Self Test and Calibration
  • Auto Lamp and Filter Selection
  • Simple Kinetics
  • Deuterium Lamp Save
  • UV Lamp Hours Indicator
  • 1500 Hours UV Lamp Guarantee

Photos

Cecil Instruments Series 1000

Cecil Instruments Series 1000

Manuals and Resources

Note: We appreciate the support of individuals and companies making their documents and resources available to our community. If you find a document or resource that does not belong please let us know by emailing info@labwrench.com

Instrument Documents

Picture of Cecil Instruments 1000 Series
Cecil Instruments are leading designers and manufacturers of HPLC systems, Ion Chromatography systems and UV / Visible Spectrophotometers.  
 
Cecil...

More about Cecil Instruments

Visit Cecil Instruments

Questions and Answers

Recent Question Activity

Ask a Question

Error code 7 (E7) on Cecil Spectrophotometer

Equipment Name: Cecil Spectrophotometer Model CE1021, 1000 series. At startup the equipment report E7 on the display. According to the operation manual the possible causes for the error is as follows: Error code    Error...

miroljub

Hi. What means an error E 36, on ce1021

Test Notifications

Test notifications

See all 9 Cecil Instruments 1000 Series Questions and Answers There are 9 Threads