Beckman Coulter DU 800 Series

Share:

DU 800 Series

The DU 800 Spectrophotometer is a PC controlled system intended for use in quantitative and qualitative analysis in biological and industrial procedures that require spectrophotometric measurements in the UV and visible region of the electromagnetic spectrum. If the instrument is used in a manner other than as described, the safety and performance of the instrument can be impaired.

The DU 800 Spectrophotometer operates in the wavelength range of 190 to 1100 nm and has a bandwidth of <= 1.8 nm. The focused micro-beam design provides a wide linear range and other specific benefits for small volumes and precious samples. Various accessories are available to address micro-volume samples and individual application requirements.
DU 800 Series
Attributes
Bandwidth 1.8 nm (from 200 to 680 nm)
Optics Single Beam
Photometric Noise < 0.0002A RMS (at 0A, average of 10 standard deviations of 10 readings at 0.05 sec intervals, at 500 nm
Scanning Speed 120, 240, 600, 1200, 2400 nm/min
Stray Light (measured using NaI at 220nm, per ASTM E387-84): < 0.05%T
Wavelength Range 190 to 1100 nm
Depth 10 in
Height 21 in
Width 27 in
Power Requirements 110 - 240V, 50/60 Hz, 200 W
Weight 81 lbs
Additional Specifications Wavelength Setability: 0.1 nm Increments
Wavelength Accuracy: Full Range + 0.5 nm
Wavelength Repeatability: Full Range + 0.2 nm
Photometric Readout: -0.300 to 3.000A or 0.0 to 200%T
Photometric Accuracy: + 0.005A (at 1A with NIST 930D solid filter at 546 nm)
Stability: < 0.003A/hr (0A, constant ambient conditions, measured for one hour at 340 nm after one hour warmup)
RMS Baseline Flatness: + 0.001A RMS (from 200 to 900 nm, at 0 A)

Features

  • Single Component Analysis
  • Kinetics/time
  • Nucleic Acid Analysis
  • RediScan™ and RediRead™
  • FixedWavelength Analysis (up to 12 wavelengths)
  • Performance Validation
  • Wavelength Scan II (peak/valley picker, point picker, derivatives, netA calculations, scatter correction and spectra in LogA)
  • 21 CFR Part 11 Compliance

Photos

DU 800 UV/Visible Spectrophotometer

DU 800 UV/Visible Spectrophotometer

Manuals and Resources

Note: We appreciate the support of individuals and companies making their documents and resources available to our community. If you find a document or resource that does not belong please let us know by emailing info@labwrench.com

Instrument Documents

Picture of Beckman Coulter DU 800 Series
Beckman Coulter, Inc. is a leading manufacturer of biomedical testing instrument systems, tests and supplies that simplify and automate laboratory processes....

More about Beckman Coulter

Visit Beckman Coulter

Questions and Answers

Recent Question Activity

Ask a Question

Beckman Coulter DU 800 not communicating via USB to PC

I have a BC DU 800 that will not communicate with the PC. I have uninstalled and reinstalled the software on the PC like a new installation. The USB port sees the device, the software is seeing the device, but they will not talk. Any suggestions?

Wavelength calibration failed

  I perform the wavelength calibration as the page # 107 manual says (both lams on , both lamps are brand new). It gives me an error message as bellow: "Wavelength calibration failed " I am wondering if anyone can help...

Godrule

My Spectrophotometer Is on and all UV AND VISIBLE light is on. Whe performing Perfomance validation its fails on Wavelength Acuracy, Resolution AND Baseline Flatness. Please asist.

See all 8 Beckman Coulter DU 800 Series Questions and Answers There are 8 Threads