PG Instruments T70 Series

Share:

T70 Series

The T70 UV Spectrophotometer is a dual beam instrument with a fixed 2nm bandpass. The wavelength range is 190 - 1100nm.The T70 has a large 'fold away' LCD display unit which allows for complete display of Spectra. The system has a large easy to use positive touch key pad and the instrument can be used a complete standalone unit.

Models

T70
Attributes

This model was added by: WrenchAdmin2

Edit Model

Bandwidth 2nm(fixed slit)
Photometric Noise ± 0.001 Abs (500 nm, 0 Abs 2 nm spectral bandwidth)
Photometric Range -0.3 ~ 3 Abs
Stray Light < 0.12%T (220 nm, Nal; 340nm, NaNo2)
Wavelength Range 190 to 1100 nm
Additional Specifications Working mode: MPU Mode/PC Mode
Software support: MPU Software Platform/Spec UV software workstation
Wavelength accuracy: ± 0.3nm (Automatic wavelength correction)
Wavelength reproducibility: 0.2 nm
Photometric mode: Transmittance, Absorbance, Energy
Photometric accuracy:
± 0.002 Abs (0 ~ 0.5 A)
± -/--4 Abs (0.5 ~ 1 A)
± 0.3% T (0 ~ 100%T)
Photometric reproducibility:
0.001 Abs (0 ~ 0.5 A)
0.002 Abs (0.5 ~ 1 A)
0.15% T (0 ~ 100%T)
Baseline flatness: ± 0.002 Abs (190 ~ 1100 nm)
Baseline stability: 0.001 Abs/h (500 nm, 0 Abs 2 nm Spectral bandwidth, 2 hr warm-up)
T70+
Attributes

This model was added by: WrenchAdmin2

Edit Model

Bandwidth 0.5,1, 2, 5nm(variable slit)
Photometric Noise ± 0.001 Abs (500 nm, 0 Abs 2 nm spectral bandwidth)
Photometric Range -0.3 ~ 3 Abs
Wavelength Range 190 to 1100 nm
Additional Specifications Working mode: MPU Mode/PC Mode
Software support: MPU Software Platform/Spec UV software workstation
Wavelength accuracy: ± 0.3nm (Automatic wavelength correction)
Wavelength reproducibility: 0.2 nm
Photometric mode: Transmittance, Absorbance, Energy
Photometric accuracy:
± 0.002 Abs (0 ~ 0.5 A)
± -/--4 Abs (0.5 ~ 1 A)
± 0.3% T (0 ~ 100%T)
Photometric reproducibility:
0.001 Abs (0 ~ 0.5 A)
0.002 Abs (0.5 ~ 1 A)
0.15% T (0 ~ 100%T)
Baseline flatness: ± 0.002 Abs (190 ~ 1100 nm)
Baseline stability: 0.001 Abs/h (500 nm, 0 Abs 2 nm Spectral bandwidth, 2 hr warm-up)

Features

  • Accurate analysis: Holographic grating greatly reducing stray light of the instrument and making the analysis more accurate.
  • Stable performances: The split beam ratio optics ensures good stability.
  • High-speed measurement: The fastest scanning speed is over 1000nm.
  • Powerful function: The main unit of the spectrophotometer can analyse for photometric measurement, quantitative measurement, spectrum scan, DNA/Protein analysis and can print data. When connected to a computer the Spec UV software adds many additional functions, such as 3D spectrum analysis, GLP laboratory protocol. It can be applied in fast pesticide remain detection, environment protection, inspection and quarantine and other fields.
  • Convenient operation: High degree of automation, the operator only needs to press keys twice when measuring ordinary samples.
  • Easily upgraded: Many optional accessories enhance the flexibility and the measurement range of the instrument.
  • Easy routine maintenance: The simple mechanical structure and modular electrical design make the routine maintenance easy.
    Original technology: The deuterium and tungsten lamps can be easily replaced and are supplied pre-aligned.
    A motorised automatic 8-cell holder is supplied as standard which is particularly useful
    for the determination of pesticides.

Photos

T70

T70

Manuals and Resources

Note: We appreciate the support of individuals and companies making their documents and resources available to our community. If you find a document or resource that does not belong please let us know by emailing info@labwrench.com

Instrument Documents

Picture of PG Instruments T70 Series
PG Instruments Limited is a British company that has for many years manufactured a comprehensive range of UV-Visible Spectrophotometers and Atomic Absorption...

More about PG Instruments

Visit PG Instruments

Questions and Answers

Recent Question Activity

Ask a Question

manual for the T70 available?

Hi, our lab has a T70 and I'm looking for a pdf copy of the manual. Does anyone have a link to that resource?  Thanks so much - Lynne

motorised 8 cell changer

Hello, Im having a problem with an T70 PG UV/VIS spectrophotomether. The motorized cell does not moves at all when the equipment is doing the initial self check.  We have to move it by ourself to finish the checking process. Sometimes...

Initialization failed

scan motor initialization, D2 lamp energy, and wavelength check failed. What shall I do?

See all 4 PG Instruments T70 Series Questions and Answers There are 4 Threads