Shimadzu SIL-20A/20AC

Share:

SIL-20A/20AC

Combining the smaller footprint of the SIL-10ADvp with the rack style and speed of the SIL-HT, the SIL-20A/20AC Autosampler provides excellent analysis efficiency through improvements in sample injection speed, accuracy, and processing capacity. Revolutionary cross-contamination countermeasures make it possible to perform analysis without carryover, even with high-sensitivity LC-MS/MS detection.  
  
SIL-20A - High-Throughput Analysis  
  
A total-volume injection type autosampler, the SIL-20A enables high-speed injection and high-throughput sample processing without wasting sample. It is designed to ensure greater stability, with improved durability attained through modifications in valves and sample loops. Multiple vial and plate types are supported, and a control vial rack isolates standards from analytical samples.  
  
SIL-20AC - Efficient Sample Cooling  
  
The SIL-20AC is equipped with a sample cooler with a built-in dehumidifier to minimize condensation problems. Samples can be maintained at a fixed temperature from 4ºC to 40ºC. Temperature equilibration is rapid and uniform, so heat or cold sensitive sample constituents remain stable.

Models

SIL-20A
Attributes

Edit Model

Cross Contamination .005 % (Chlorhexidine)
Optional Injection Range 1 to 2000 ml
Standard Injection Range .1 to 100 ml
Number of Samples 105 1.5 mL vials
Depth 500 mm
Height 415 mm
Width 260 mm
Power Requirements AC110-120V, 100VA, 50/60Hz
Weight 27 kg
Additional Specifications Injection Method:
Metering pump, variable volume, no sample loss
Sample Capacity:
Control Rack: 10 1.5 mL vials
Optional Sample Racks for:
175 1 mL vials
50 4 mL vials
2 96 well MTP (std or deep well)
2 384 well MTP
Injection Volume Accuracy:
+ 1% (50ΜL, n=10)
Injection Volume Precision:
RSD < 0.3% (10ΜL)
Sample Aspiration Rate:
0.1-15 ΜL/sec (0.1ΜL steps)
Rinsing Rate:
1-35 ΜL/sec (1ΜL steps)
Rinsing Options:
Before Aspiration, After Aspiration, Before and After Aspiration, or None
(Optional rinse pump for multiple solvent wash available)
pH Range:
1 to 14
Operating Temperature:
4oC to 35oC (non-condensing atmosphere)
SIL-20AC
Attributes

Edit Model

Depth 500 mm
Height 415 mm
Width 260 mm
Weight 30 kg
Additional Specifications Injection Method:
Metering pump, variable volume, no sample loss
Injection Volume:
0.1 to 100 ΜL (std), 1 to 2,000 ΜL (opt)
Sample Capacity:
70 1.5 mL vials (incl. std.);
Control Rack: 10 1.5 mL vials
Optional Sample Racks for:
175 1 mL vials
50 4 mL vials
2 96 well MTP (std or deep well)
2 384 well MTP
Injection Volume Accuracy:
+ 1% (50ΜL, n=10)
Injection Volume Precision:
RSD < 0.3% (10ΜL)
Carryover:
0.005% (Chlorhexidine)
Sample Aspiration Rate:
0.1-15 ΜL/sec (0.1ΜL steps)
Rinsing Rate:
1-35 ΜL/sec (1ΜL steps)
Rinsing Options:
Before Aspiration, After Aspiration, Before and After Aspiration, or None
(Optional rinse pump for multiple solvent wash available)
Temperature Control:
4o-40oC w/ built-in dehumidifier
pH Range:
1 to 14
Operating Temperature:
4oC to 35oC (non-condensing atmosphere)
Power Requirements:
AC110-120V, 300VA, 50/60Hz

Features

Near-Zero Carryover   
Adsorption of sample constituents has been reduced to an absolute minimum. Enhancements include:  
      
  • Special needle treatment (patent pending) - masks active sites and minimizes ionic interactions  
  • One piece injection port/needle seal at valve center - eliminates sample transfer tubing  
  • Redesigned needle seal to minimize surface area  
  • Low-adsorption valve materials  
  
These advancements combine to eliminate cross contamination - even with difficult basic and hydrophobic compounds. For especially difficult samples, dual-solvent rinsing is available.   
  
High Speed/Throughput - 10s for a 10 µL injection  
Ultra-high-speed sample processing - 10 seconds for a 10 µL injection - is now possible with the SIL-20A/20AC. Used together with a high-speed separation column makes an analysis cycle time of less than one minute a reality.  
  
Precise Sample Injection  
Greater injection volume accuracy has been attained by incorporating a high-performance precision metering pump. Injection volumes of less than 1 µL are possible, and the direct injection method means valuable samples are never wasted.  
  
Optional Rack Changer  
Significantly Increases Sample Capacity  
  
With increased speed, capacity becomes a limiting factor for unattended operation. The optional rack changer automatically loads microplates into the SIL-20A/AC sample compartment. The 12-plate capacity ensures convenient processing of over 1000 samples. The "C" model changer has the same plate capacity and incorporates a 4ºC to 40ºC cooling function, matching that of the SIL-20AC.

Photos

SIL-20A/20AC

SIL-20A/20AC

Manuals and Resources

Note: We appreciate the support of individuals and companies making their documents and resources available to our community. If you find a document or resource that does not belong please let us know by emailing info@labwrench.com

Instrument Documents

Picture of Shimadzu SIL-20A/20AC
Shimadzu Scientific Instruments (SSI) is the American subsidiary of Shimadzu Corporation, headquartered in Kyoto, Japan. Founded in 1875, Shimadzu has...

More about Shimadzu

Visit Shimadzu

Questions and Answers

Recent Question Activity

Ask a Question

Needle Alignment Procedure for SIL-20AC HT

Does anyone have the service manual with needle to needle seal alignment procedure? Thank You, Scott

There are 1 Threads

Reviews

Member Reviews

There are no threads in Shimadzu SIL-20A/20AC Reviews at this time.
Add a Review

Buy it Used!

Used Shimadzu SIL-20A/20AC Offerings on LabX

Shimadzu Prominence 20 HPLC UFLC System Including Software

Price: $19,990.00

Condition:

Configuration: As Shown

SIL-20AC

Price:

SIL-20A HT

Price:

SIL-20A HT

Price:

Find more results for "Shimadzu SIL-20A/20AC" models on LabX