Advertisement

Analytik Jena multi N/C®

Share:

multi N/C®

The TOC/TNb analyzers of the multi N/C® series set new standards!

The parameters TOC, NPOC, POC, TC, TIC and TNb can be easily measured precisely, fast and reliably. multi N/C® is the ideal system for routine applications, such as the examination of drinking and waste water, or special applications, such as pharmaceutical purification validation.

The modular versatility and comprehensive range of accessories guarantee for every user an individual adaptation to the respective requirements. The fast readiness for operation, easy handling and long-lasting components make every model within the series to an ideal partner during daily practice.

Models

multi N/C® 2100
Attributes

Edit Model

Measurement Range 1 to 30000 mg/L
Detection Limit 200 µg/L
Measurement Time approx. 3- 5 min for each parameter
NDIR Yes
Sample Volume Range 50 to 500 μL variable
Portable No
Depth 550 mm
Height 464 mm
Width 513 mm
Power Requirements 115/230 V; 50/60 Hz; 10A/ 6.3A
Peak Power 1000 VA
Weight 30 kg
Additional Specifications Digestion mode: High temperature combustion up to 950°C 
Detection of CO2: Focus Radiation NDIR detector® 
Control by: PC 
Parameters: TC, TIC, TOC, NPOC, NPOC plus, POC (indirect measurement) 
Optional parameters: TN (ChD or CLD) 
Measurement accuracy (reproducibility): CV 2-3% 
Sample injection: direct injection 
Special features: Auto-protection, Self Check System 
TIC elimination for NPOC mode: acidification and purging + TIC control measurement 
Automatic system tightness control: no 
Required gas: synthetic air or oxygen 
Gas consumption: approx. 1800 l/month (calculated for 8 hours/day x 5 days/week operation) 
Upgrade for solid samples: Double furnace solid module or HT 1300 
Sampler options: APG 60 
Included Software: PC software multiWin®
multi N/C® 2100S
Attributes

Edit Model

Measurement Range 0 to 30000 mg/L
Detection Limit 50 µg/L
Measurement Time approx. 3- 5 min for each parameter
NDIR Yes
Sample Volume Range 50 to 500 μL variable
Portable No
Depth 550 mm
Height 464 mm
Width 513 mm
Power Requirements 115/230 V; 50/60 Hz; 10A/ 6.3A
Peak Power 1000 VA
Weight 30 kg
Additional Specifications Digestion mode: High temperature combustion up to 950°C 
Detection of CO2: Focus Radiation NDIR detector® 
Control by: PC or stand alone option 
Parameters: TC, TIC, TOC, NPOC, NPOC plus, POC (indirect measurement) 
Optional parameters: TN (ChD or CLD) 
Measurement accuracy (reproducibility): CV 1-2% 
Sample injection: Septum-free direct injection 
Injection volume: 50 – 500µl variable 
Special features: VITA® Flow Management System, Auto-protection,  
Self Check System (SCS), Easy Cal® 
TIC elimination for NPOC mode: acidification and purging + TIC control measurement 
Automatic system tightness control: yes 
Required gas: synthetic air or oxygen 
Gas consumption: approx. 1800 l/month (calculated for 8 hours/day x 5 days/week operation) 
Upgrade for solid samples: Double furnace solid module or HT 1300 
Sampler options: APG 60 
Included Software: PC software multiWin® 
Software Options: Stand alone software (included in SA-Option)
multi N/C® 3100
Attributes

Edit Model

Measurement Range 1 to 30000 mg/L
Detection Limit 4 µg/L
Measurement Time approx. 3- 5 min for each parameter
NDIR Yes
Sample Volume Range 100 to 1000 μL variable
Portable No
Depth 550 mm
Height 464 mm
Width 513 mm
Power Requirements 115/230 V; 50/60 Hz; 10A/ 6.3A
Peak Power 1000 mW
Weight 30 kg
Additional Specifications Digestion mode: High temperature combustion up to 950°C 
Detection of CO2: Focus Radiation NDIR detector® 
Control by: PC or stand alone option 
Parameters: TC, TIC, TOC, NPOC, NPOC plus, POC (indirect measurement) 
Optional parameters: TN (ChD or CLD), POC direct 
Measurement accuracy (reproducibility): CV 1-2% 
Sample injection: Automatic flow injection with intelligent rinsing 
Special features: VITA® Flow Management System, Auto-protection,  
Self Check System (SCS), Easy Cal® 
TIC elimination for NPOC mode: acidification and purging + TIC control measurement 
Automatic system tightness control: yes 
Automatic dilution: Dilution factor 2 up to 100 
Required gas: synthetic air or oxygen 
Gas consumption: approx. 2200 l/month (calculated for 8 hours/day x 5 days/week operation) 
Upgrade for solid samples: HT 1300 
Sampler options: APG 10, APG 21, APG 64, EPA 
Included Software: PC software multiWin® 
Software Options: Stand alone software (included in SA-Option)
multi N/C® UV HS
Attributes

Edit Model

Measurement Range 1 to 10000 mg/L
Detection Limit 2 µg/L
Measurement Time approx. 3- 5 min for each parameter
NDIR Yes
Sample Volume Range 50 to 20000 μL variable
Portable No
Depth 550 mm
Height 464 mm
Width 513 mm
Power Requirements 115/230 V; 50/60 Hz; 10A/ 6.3A
Peak Power 1000 VA
Weight 30 kg
Additional Specifications Digestion mode: Wet chemical oxidation by UV (254nm+185 nm) / persulfate 
Detection of CO2: Focus Radiation NDIR detector® 
Control by: PC 
Parameters: TC, TIC, TOC, NPOC, NPOC plus, POC (indirect measurement) 
Measurement accuracy (reproducibility): CV 1-2% 
Sample injection: Automatic flow injection 
Special features: VITA® Flow Management System, Auto-protection,  
Self Check System (SCS), Easy Cal® 
TIC elimination for NPOC mode: acidification and purging + TIC control measurement 
Automatic system tightness control: yes 
Required gas: nitrogen 
Gas consumption: approx. 2200 l/month (calculated for 8 hours/day x 5 days/week operation) 
Upgrade for solid samples: HT 1300 
Sampler options: APG 10, APG 21, APG 49, EPA sampler 
Included Software: PC software multiWin®
multi N/C® pharma UV
Attributes

Edit Model

Measurement Range 0 to 10000 mg/L
Detection Limit 2 µg/L
Measurement Time approx. 3- 5 min for each parameter
NDIR Yes
Sample Volume Range 50 to 20000 μL variable
Portable No
Depth 550 mm
Height 464 mm
Width 513 mm
Power Requirements 115/230 V; 50/60 Hz; 10A/ 6.3A
Peak Power 1000 VA
Weight 30 kg
Additional Specifications Digestion mode: Wet chemical oxidation by UV (254nm+185 nm) / persulfate 
Detection of CO2: Focus Radiation NDIR detector® 
Control by: PC 
Parameters: TC, TIC, TOC, NPOC, NPOC plus, POC (indirect measurement) 
Measurement accuracy (reproducibility): CV 1-2% 
Sample injection: Automatic flow injection 
Special features: VITA® Flow Management System, Auto-protection,  
Self Check System (SCS), Easy Cal® 
TIC elimination for NPOC mode: acidification and purging + TIC control measurement 
Automatic system tightness control: yes 
Required gas: nitrogen 
Gas consumption: approx. 2200 l/month (calculated for 8 hours/day x 5 days/week operation) 
Sampler options: APG 10, APG 21, APG 49, EPA sampler 
Included Software: Pharma PC software multiWin®
multi N/C® pharma HT
Attributes

Edit Model

Measurement Range 1 to 10000 mg/L
Detection Limit 4 µg/L
Measurement Time approx. 3- 5 min for each parameter
NDIR Yes
Sample Volume Range 50 to 30000 μL variable
Portable No
Depth 550 mm
Height 464 mm
Width 513 mm
Power Requirements 115/230 V; 50/60 Hz; 10A/ 6.3A
Peak Power 1000 VA
Weight 30 kg
Additional Specifications Digestion mode: High temperature combustion up to 950°C 
Detection of CO2: Focus Radiation NDIR detector® 
Control by: PC 
Parameters: TC, TIC, TOC, NPOC, NPOC plus, POC (indirect measurement) 
Optional parameters: TN (CLD) 
Measurement accuracy (reproducibility): CV 1-2% 
Sample injection: Automatic flow injection 
Special features: VITA® Flow Management System, Auto-protection,  
Self Check System (SCS), Easy Cal® 
TIC elimination for NPOC mode: acidification and purging + TIC control measurement 
Automatic system tightness control: yes 
Required gas: oxygen 
Gas consumption: approx. 1400 l/month (calculated for 8 hours/day x 5 days/week operation) 
Upgrade for solid samples: Swab test module 
Sampler options: APG 10, APG 21, APG 49, EPA sampler 
Included Software: Pharma PC software multiWin®

Features

  • Large measuring range – even without sample dilution: precise detection due
    to the high quality Focus Radiation NDIR Detector® (FR NDIR Detector) for TOC, Chemoluminescence detector (CLD) or solid state chemical detector (ChD) for TNb
  • VITA® Flow Management System: for stable equipment performance and highly reproducible analysis results
  • Easy Cal®: easiest calibration with only one standard for the most varied applications, including longterm stability
  • Auto-protection: effective measuring gas purification and monitoring protect high quality system components
  • Reliable oxidation: high temperature incineration (up to 950ºC) or High power UV reactor
  • Variable injection technologies: valve-free direct injection or flow injection
  • Variable injection volumes
  • TC, TOC, TIC, NPOC and NPOC plus, POC, TNb: multiple parameter detection
  • Suitable for the simultaneous TNb detection
  • Solids analysis at the highest level: Double furnace technology — the ideal combination for water and solid samples without additional furnace
  • Self Check System (SCS) for optimum operational and analytic safety
  • Automation and modular principle: multitude of analyzers and accessories tailored for your application

Photos

  • multi N/C® Series
  • Easy Cal®
  • Focus Radiation NDIR-Detector®
  • High Power UV reactor
  • VITA®  Flow Management System
multi N/C® Series

multi N/C® Series

Manuals and Resources

Note: We appreciate the support of individuals and companies making their documents and resources available to our community. If you find a document or resource that does not belong please let us know by emailing info@labwrench.com

Instrument Documents

Picture of Analytik Jena multi N/C®
Analytik Jena is a complete system provider of analytical measuring technology. The portfolio extends from traditional analysis technology to bioanalytical...

More about Analytik Jena

Visit Analytik Jena

Questions and Answers

Recent Question Activity

Ask a Question

NPOC method

Hi, I want to ask why we need to acidify and purge the sample in the NPOC method of the TOC analyzer to determine Total organic carbon. Thanks you Khalid

service engineer

let me the specification of the thermocouple (heater sensor for furnace) analytik jena TOC model multi n/c 3100?

Analysis: Determing Nitrogen content with the multi AE5000

I have a question. I am determing Nitrogen content with the multi AE5000.     Previous lab results obtained data that are 10 times higher then our data.    The calibration is done accoriding to the manual standards. ...

There are 3 Threads

Reviews

Member Reviews

There are no threads in Analytik Jena multi N/C® Reviews at this time.
Add a Review