Advanced Instruments - Fiske® 210

Manufactured by  Advanced Instruments
Follow this Equipment

Fiske Associates Osmometer, Model 210, for Osmolality Measurement

The Fiske Associates Model 210 Micro Osmometer determines the osmolality of solutions using freezing point depression (FPD). Ideal for serum, urine or other bodily fluids in clinical applications; and media or formulations in pharmaceutical, biotech or industrial applications.  
  
Automated, single sample, designed to process a 20 mL sample, and 90-second test time. The Model 210 provides the best accuracy and precision in the market today.
Active Questions & AnswersAsk a Question

There are no current Discussions

Osmometers Service ProvidersView All (5)

Request Support
Features of Fiske® 210
  • 20 µL sample size is ideal for applications where the sample quantity is limited.  
  • Unparalleled ease of use  
  • Solid state cooling  
  • Internal diagnostics  
  • Ergonomic, compact design  
  • On-board statistical analysis  
  • Results recall up to the last 30 tests  
  • Multi-language operation  
  • Automated calibration  
  • Hot-Line® Technical Service available 24/7
General Specifications
Depth 9.5 in
Height12 in
Width 12 in
Power Requirements 100 to 240 VAC (50/60 Hz)
Peak Power60 W
Weight13.6 lbs
Additional Specifications

Sample Capacity: Single sample  
Readout: 2 line by 24-character vacuum fluorescent display  
Units: mOsm/kg H2O  
Range: 0 to 2000 mOsm/kg H2O  
Resolution: 1 mOsm/kg H2O  
Communications: DTE EIA-232/V.24 (RS-232); parallel printer port and optional barcode scanner  
  
Performance at Reference Conditions  
Linearity: Less than ±1% from a straight line between 0 and 2000 mOsm/kg H2O  
Repeatability: 0 to 400 mOsm: ±2 mOsm/kg (1 S.D.) 400 to 2000 mOsm: ±0.5% (1 S.D.)  
Drift: Less than 1 mOsm/kg H2O per month  
  
Performance Over Operating Conditions  
Temperature Effects: Less than 1 mOsm/kg H2O for every 5°C (9°F) ambient temperature change  
Operating Conditions Temperature: 18°C to 35°C (64°F to 95°F)  
Humidity: 5 to 80% relative humidity (non-condensing)  
Storage Temp.: -40°C to +45°C (-40°F to +113°F)  
Start-up Time: 30 seconds from power-on  
Test Time: 90 seconds

Looking for

Advanced Instruments - Fiske® 210?

Shop on LabX.com