Setaram Instrumentation - microDSC3 evo
Manufactured by Setaram Instrumentation
High-sensitivity DSC and isothermal microcalorimetry (-20 to 120°C) with in-situ mixing possibilities
ΜDSC3 evo is the lastest generation of the most popular microcalorimeter of Setaram. Its high sensitivity and versatility makes this instrument a star in many industrial and academic research laboratories all around the world.
Active Questions & AnswersAsk a Question
There are no current Discussions
Need Equipment Support?
Documents & ManualsView All Documents
Features of microDSC3 evo
- Wide operating temperature range (from -20 °C to 120 °C): Cooling of the calorimeter does not require an external source.
- Very low detection limit: The µDSC3 evo detection limit is significantly lower (by a factor of about ten) than that of traditional DSCs. Calorimetric signals of less than one Microwatt can be detected. The very high sensitivity of the µDSC3 evo enables transition studies to be performed on small quantities of products and at very low programming speeds.
- Use in isothermal or temperature programming (DSC) mode.
- Possibility of studying samples in all forms: liquid, gel, powder, solid.
- Large range of experimental cells: including specific cells for studying reactions, providing numerous application possibilities.
- Ease of operation: dedicated market leading CALISTO software that is not only intuitive but powerful enough to perform all typical experiments and data treatment.
General Specifications
Cooling Rate | 0.001 to 1.2 °C/min |
Depth | 22.8 in |
DSC Type | Heat Flux DSC |
Height | 15.7 in |
Temperature Range | -20 to 120 °C |
Width | 20.9 in |
Heating Rate | 0.001 to 1.2 °C/min |
Power Requirements | 230 V - 50/60 Hz |
Weight | 82.5 lbs |
Additional Specifications
Temperature range:
-20 °C to 120 °C (Cooling under 0 °C requires the use of an auxiliary thermostat)
Programmable temperature scanning rate (heating and cooling):
0.001 to 1.2 °C.min-1-1
RMS Noise:
0.2 µW
Resolution:
0.02 µW / 0.002 µW
Cells ("batch" or circulation):
1 ml - made of Hastelloy C - Removable
Pressure (measured & controlled):
400 bar / 5800 psi, requires the use of High Pressure cells and gas panel