An-easy-to-use Differential Scanning Calorimeter DSC (-170 to 700°C)

DSC131 evo: for testing, QC and academic thermal analysis laboratories. An instrument designed from the ground up to be robust, high performance and above all user friendly featuring market leading intuitive CALISTO software and low operating costs.
Active Questions & AnswersAsk a Question

Recent Questions & Answers

avatar placemark

Asked bymirosero

avatar placemark

Asked byjlaforce

Differential Thermal Analysis Service ProvidersView All (8)

Request Support
Features of DSC131 evo

Robustness: the DSC131 evo features a highly robust sensor that, if needed, can readily be changed by the user in less than 30 minutes. No other DSC offers such flexibility.    

Ease of operation: dedicated market leading CALISTO software that is not only intuitive but powerful enough to perform every typical experiment and data treatment.    

Flexible and powerful: from -170 to 700 °C, with rapid heating and cooling performance.    

Large range of crucibles: in addition to the regular 30 µl crucibles, crucibles with a capacity of 100 µl are used for analyzing heterogeneous samples and for optimizing Cp (heat capacity) measurements.    

High Pressure Crucibles: The High Pressure crucibles deliver unmatched High Pressure resistance (up to 500 bars / 7 250 psi, 600 °C) while the DSC sensor itself remains at atmospheric pressure.

General Specifications
Cooling Rate0.01 to 100 °C/min
Depth 22.8 in
DSC TypeHeat Flux DSC
Height15.7 in
Temperature Range28 to 700 °C
Width 20.9 in
Heating Rate0.01 to 100 °C/min
Power Requirements 230 V - 50/60 Hz.
Weight82.5 lbs
Additional Specifications

Temperature range:     
Ambient to 700 °C    
With cooling accessories:     
-170 °C to 500 °C (Liquid Nitrogen Accessory)    
-70 °C to 400 °C (Cryothermostat)    
Programmable temperature scanning rate(heating and cooling):     
0.01 to 100 °C.min-1    
Cooling time:     
12 min (500 °C to 100 °C) Air    
5 min (100 °C to 0 °C) Cryothermostat    
6 min (200 °C to 25 °C) Liquid Nitrogen Accessory    
12 min (25 °C to -100 °C) Liquid Nitrogen Accessory    
RMS Noise (200°C):     
1.5 µW    
Resolution:     
0.8 µW    
Time constant:     
3 s    
Gases:     
included gas switch from gas A to gas B    
Crucibles:     
30 µl, 100 µl    
aluminum, alumina, incoloy, etc.    
Pressure (non controlled):     
High pressure crucible (up to 500 bars / 7255 psi at 600 °C)