Setaram Instrumentation - LABSYS evo

Manufactured by  Setaram Instrumentation
Follow this Equipment

TGA, DTA, DSC and Simultaneous TGA-DTA, TGA-DSC from ambient to 1600°C

LABSYS evo offers TGA with simultaneous DTA or DSC and Cp measurement within the temperature range of ambient to 1 600 °C. The LABSYS evo features a NEW high performance top loading balance and a unique 3D high-precision (Calvet type) Cp rod. Thanks to its unique ergonomic design and market leading Calisto software the LABSYS evo offers you the most easy-to-use Thermal Analysis system with uncompromised performance.
Active Questions & AnswersAsk a Question

Recent Questions & Answers

avatar placemark

Asked byMuddasser

Need Equipment Support?

Request Support
Features of LABSYS evo

Simplicity of use: thanks to both an ergonomic design and the dedicated market leading Calisto software, LABSYS evo is user friendly and intuitive to operate    

Wide temperature range: TGA with simultaneous DTA or DSC and Cp measurement can be achieved for temperature up to 1 600 °C    

High Precision Cp Measurement: the unique 3D (Calvet type) multiple thermocouples design enables large mass to be measured and to perform a Cp within 2 % , throughout the whole temperature range (ambient – 1 600 °C)    

New top loading balance: the unmatched stability, reproducibility and accuracy of this new balance offer unparalleled TGA data

General Specifications
Cooling Rate0.01 to 100 °C/min
Depth 22.8 in
DSC TypeHeat Flux DSC
Height22 in (closed)
Temperature Range28 to 1600 °C
Width 20.8 in
Heating Rate0.01 to 100 °C/min
Power Requirements 230 V - 50/60 Hz
Weight121 lbs
Additional Specifications

Temperature Range:     
Ambient to 1 600 °C    
Isothermal temperature accuracy:     
+/- 1 °C    
Isothermal temperature precision:     
+/- 0.04 °C    
Programmable temperature scanning rate (heating and cooling:     
0.01 to 100 °C.min-1    
Furnace cooling:     
32 min (1600°C to 50°C)    
Maximum balance capacity:     
20 g    
Weight range:     
+/- 1 000 mg - +/- 200 mg    
Weighing precision:     
+/- 0.01 %    
TG resolution:     
0.2 Μg - 0.02 Μg    
Baseline dynamic drift (one hour):     
10 Μg    
TG reproductibility up to 1 200 °C:     
< 50 Μg    
DSC rod - Resolution:     
0.4 µW / 10 ΜW    
3D Cp rod - Cp Accuracy:     
< 2 %    
Vacuum:     
< 10-1 mbar    
Evolved gas:     
simultaneous MS, FTIR couplings (option)    
Gases:     
Two inlets for gas scanning (inert or reactive) OR 3 carrier gases (option) (MFC from 4 to 200 ml/min) + 1 auxiliary or reactive gas (MFC from 0.3 to 16 ml/min)

Videos
  • Userfriendly Thermal Analysis Platform LABSYS evo with Auto Sampler

Looking for New or Used Equipment?

Shop on LabX