Biochrom - Biowave DNA

Manufactured by  Biochrom
Follow this Equipment

WPA Biowave DNA Life Science Spectrophotometer

The Biowave DNA has been specifically designed for life science applications and is a powerful tool for the laboratory that requires a dedicated instrument for the determination of nucleic acid purity and concentration, protein concentrations or cell density measurements.     
     
The system utilises Novel Optics for high energy throughput, a Xenon light source for long lamp lifetimes together with simple selection software and large graphical display for ease of use and data interpretation. The stored methods include   DNA, RNA and oligonucleotide calculations, protein assays such as direct UV measurement, BCA, Biuret, Bradford and Lowry and cell density measurement.   Unlike many dedicated life science instruments the Biowave DNA can also measure Absorbance or concentration at any wavelength so there is complete flexibility for future applications.     
     
For added convenience it is possible to display a scan of the nucleic acid profile which is particularly useful for RNA samples where impurities may be present in the 230 nm region, yet not have an adverse effect on the 260/280 Absorbance ratio. The system is compatible with both Quartz and disposable low volume UV cuvettes.      
     
Results may be printed to an optional integrated high quality graphical printer for permanent record or exported via USB cable connection or SD Card to a suitable PC running optional Print Via Computer (PVC) software for advanced reporting or data storage.
Active Questions & AnswersAsk a Question

There are no current Discussions

UV/VIS Spectrophotometers Service ProvidersView All (32)

Request Support
Features of Biowave DNA
  • Split Beam (RBC) UV/Vis Spectrophotometer 190-900nm (5nm bandwidth)     
  • Pre-programmed methods for DNA, RNA, proteins and cell density     
  • Nucleic acid scans for purity checking (scan 220-330nm)     
  • Low volume capability with 10µL min. quartz cells     
  • Novel optics with a Xenon source for long lamp lifetime and low running costs     
  • Fleible data export optional accessories include Export to PC software, built in printer or SD card Method/Data storage.
General Specifications
Bandwidth5nm
Depth 15.3 in
Height3.9 in
Width 10.2 in
Power Requirements 90-250 V AC, 50/60Hz
Peak Power30 VA
Photometric Range-0.300 to 2.500A, 0 to 199%T
Weight9.9 lbs
Wavelength Range190 to 900 nm
Light SourceXenon (PTR)
OpticsSplit Beam
Stray Light0.5%T at 220 & 340nm
Additional Specifications

Wavelength Accuracy (+/-): 2nm      
Wavelength Reproducibility (+/-): 1nm      
Optical System: Dual channel Monochromator      
Detector: 1024 CCD array      
Photometric Accuracy: +/- 0.003A at 0-0.5A at 546nm      
Stability: 0.001A/h at 340nm      
Spectral Scanning: Yes (220-320nm)      
Measurement Type: Fixed wavelength (Abs/%T), Abs ratio and Abs difference in nucleic acid mode only, concentration (factor), Life Science Methods, cell density      
Cell Compatibility: Cuvette 10mm, Microcell* , Ultra Microcell*      
Other Sample Containers: n/a      
Cell Holder: Single Cell      
Accessory Options: None      
Temperature Control: None      
Method Storage: 9      
Regulatory Compliance: CE      
Printer: Thermal built in*, External via Export*      
Data Output: Print*, Store*, Export to PC*      
Display: Graphical LCD      
Export/ Data Storage: SD card*, Bluetooth*, USB cable*      
Data Interface: USB      
Languages: English