Zoex Corporation - FasTOF™

Manufactured by  Zoex Corporation
Follow this Equipment

GC x GC x HiResTOFMS

Zoex Corporation introduces FasTOF™, a high resolution time-of-flight mass spectrometer detector for GC x GC. A pulsed mass calibration system compensates for temporal drift in mass calibration parameters. GC Image Software fully supports high resolution mass spectrometric analyses, including exact mass measurements and elemental composition determinations.
Active Questions & AnswersAsk a Question

There are no current Discussions

GC-TOF MS Service ProvidersView All (1)

Request Support
Documents & Manuals

There are no Documents or Manuals available.

Features of FasTOF™

FasTOF™ Mass Analyzer Assembly  The mass analyzer is a classical reflectron with a 1.2 meter ion flight path. Ions are detected on a high speed ADC channel, using dual Chevron micro channel plates 

FasTOF™ Ion Source  The FasTOF™ Ion Source provides adjustable 0 eV to 100 eV electron impact ionization. Ions are orthogonally extracted into the flight tube. Ion source and analyzer sections are differentially pumped by a multi-stage turbo molecular and oil-free mechanical pumping system. 

GC Image  GC Image software provides full service analysis of high resolution mass spectral data, including exact mass measurement and elemental composition determinations. High resolution mass spectrometry capabilities are compatible with GC Image template system for analysis of GC x GC chromatograms, CLIC expressions for rapid identification of compound classes and targets, and with GC Project platform, which supports automatic data analysis from autosampler sequences. 

FasTOF™ supports a full range of practical GC x GC x HiResTOFMS applications, including metabolomics, fuels, flavors and fragrances, and environmental analyses. 

Mass Resolution  The FasTOF™ provides tunable mass resolution of up to 5,000. Routine operating resolution is typically 4,000. 

Mass Accuracy  The FasTOF™ provides mass accuracy less than or equal to +/-0.002 Th (amu) across the mass calibration spectrum. Typical mass accuracy is +/-0.0007 Th. Mass calibrant is pulsed into the ion source during the dead band of the GC x GC secondary column, resulting in a new mass calibration every modulation period. This results in multi-parameter drift compensation, i.e., "set and forget" mass calibration stability. 

Mass Calibration  FasTOF™ employs a new and unique pulsed calibration system. Mass calibrant is pulsed into the mass spectrometer ion source, in synchrony with the GC x GC secondary column dead band. This permits the mass calibration to be updated every few seconds, eliminating drift in calibration parameters. 

Sensitivity  The FasTOF™ provides S/LOD ratio of 50:1 from 1.0 pg of OFN delivered to the GC Column by splitless injection. S/N (Noise, N, defined as 1.4 times the noise standard deviation) is typically greater than 300:1 from 1.0 pg of octafluoronaphthalene (OFN).

General Specifications

There are no General Specifications available.