The ELISA STARlet is a complete solution for your non-IVD ELISA assays. The system combines a high precision pipetting...

The ELISA STARlet is a complete solution for your non-IVD ELISA assays. The system combines a high precision pipetting unit ML STARlet with proven FAME incubators and new washer and photometer units.         
         
The ELISA STARlet is equipped with 8 independent channels that are assymetrically spreadable, Monitored Air Displacement technology for precise pipetting and dual (capacitative and pressure) liquid level detection for polar and non polar liquids.         
         
The system includes all necessary functions for your ELISA assay: barcode reading, plate handling, plate washing, incubation and photometer - all in a single enclosure. Based on the successful STAR line, the system offers a wide range of optional accessories including needles, washstations and CO-RE compatible manipulation tools, as well as plate, tube, reagent and custom labware carriers.         
         
The VELISA software optimizes your assays with its Dynamic Scheduler for flexible scheduling of available resources on the system. The resource management on the ELISA STARlet can optimize a standard ELISA assay to process 10 plates in 7 hours.
Active Questions & AnswersAsk a Question

There are no current Discussions

Liquid Handling Service ProvidersView All (9)

Request Support
Features

There are no Features available.

General Specifications
Depth 1005 mm
Height945.5 mm
Width 2220 mm
Power Requirements 115 V ~ / 230 V ~ 50 Hz / 60 Hz
Weight250 kg
Additional Specifications

Intended Use:      
Complete processing of ELISA assays including pipetting of the assay plates, wash cycles, incubation, read out of the assay     
plates and creation of report files in a non IVD environment.     
Throughput:     
For a standard ELISA Assay: 10 assay plates in 7 hours.     
Recommended PC:      
Pentium 4, 1024 MB RAM, 80 GB harddisk, Microsoft WIndows XP Professional, 6 USB circuit points, 2 RS232 circuit points     
Communication:      
Pipetting Unit/Washer/Reader: USB, Incubator: RS232     
     
Pipetting Unit     
Work Area Dimensions:      
Width: 675 mm     
Height: 195 mm     
Depth: 465 mm     
Deck Capacity:      
30 tracks (T) allow combinations of: maximum of 30 tube carriers (1 T) holding 24 or 32 tubes per carrier     
maximum of 5 carriers (6 T) holding 5 tip racks or 5 plate positions per carrier     
Pipetting Head:      
8 independently controlled air displacement pipetting channels     
Positional Accuracy:      
X-Y-Z positional accuracy of 0.1mm     
Tip Sizes:      
Low volume: 10Μl     
Standard volume: 300Μl      
High volume: 1000Μl     
Liquid Level Detection:     
Capacitive liquid level detection (cLLD) and pressure (pLLD) on aspiration, cLLD on dispense, minimum volume 10Μl, depending on     
container type     
Throughput:      
To fill one 96-well microtiter plate with 100Μl samples (new tips for each sample): 320s. Aliquot reagent to a 96-well microtiter     
plate (<90Μl per well): 60s     
Labware:      
all SBS standard plate types and most commercially available tube types     
Carriers:      
for all standard labware formats and according to customer requirements