PerkinElmer - Clarus SQ 8

Manufactured by  PerkinElmer
Follow this Equipment

Sensitivity. Stability. Versatility.

Introducing the new Clarus® SQ 8 Gas Chromatograph/Mass Spectrometer (GC/MS). This world-class system delivers uncompromised performance with consistent ultra-trace detection limits consistenly and reliably – time after time.           
           
The revolutionary Clarus SQ 8 GC/MS is engineered around the industry’s most sensitive, yet durable Clarifi™ detector and a unique SMARTsource™ (Simplified Maintenance And Removal Technology) that is easy to remove and clean – without tools. The Clarifi detector provides you with the flexibility to choose your ideal level of sensitivity and dynamic range, eliminates background noise and maximizes analyte signals. SMARTsource enables you to virtually reconfigure between EI and CI – simply and quickly.           
           
Clarus SQ 8 GC/MS features the fastest conventional GC oven available – heating up or cooling down more rapidly than other competitive ovens. Plus, the system is driven by our sample-centric TurboMass™ software for ease-of-use from data collection to evaluation and reporting. Combine the Clarus SQ 8 GC/MS with our market-leading TurboMatrix™ sample handling, user-friendly software and world-class service for an integrated, complete analytical solution from a single source.
Active Questions & AnswersAsk a Question

GC-Single Quad/MSD Service ProvidersView All (1)

Request Support
Features of Clarus SQ 8
               
  • World’s most sensitive detector delivers exceptional sensitivity           
  • Easiest source to access and maintain (EI and CI)           
  • Universal Programmable Split/Splitless (PSS) Injector           
  • Headspace and Thermal Desorption Sample Introduction Systems           
  • Marathon™ Filament delivers long life even under the most difficult chromatography conditions           
  • TurboMass GC/MS software delivers easy environmental compliance           
  • Fastest heat-up and cool-down oven in conventional GC           
  • Choose from: Clarus SQ 8 S, Clarus SQ 8 T, Clarus SQ 8 C           
General Specifications
Depth 32 in Clarus SQ 8 T GC/MS (with Autosampler)
DetectorClarifiTM detector—long-life high sensitivity electron multiplier detector
Height33 in Clarus SQ 8 T GC/MS (with Autosampler)
Temperature Range10 to 30 °C
Mass Range1 to 1200
Width 39 in Clarus SQ 8 T GC/MS (with Autosampler)
Power Requirements 120 VAC +– 10% @ 50/60 Hz +– 1% 1,000 VA, 230 VAC +– 10% @ 50/60 Hz +– 1% 1,000 VA
Additional Specifications

[u]Hardware[/u]         
Ion Source: PerkinElmer SMARTsourceTM—Simplified Maintenance and Removal Technology (EI and CI compatible)         
Filament: MarathonTM—designed for long life         
Analyzer: Quadrupole with pre-filter         
Mass Stability: +–0.1 m/z over 48 hours         
EI Voltage: 10 – 100 eV         
Vacuum Pumps: T and C models: 255 L/sec (nitrogen) air-cooled turbomolecular pump (230 L/sec helium pumping capacity)         
S model: 75 L/sec (nitrogen) air-cooled turbomolecular pump (59 L/sec helium pumping capacity)         
Vacuum Gauge: Single wide-range gauge (standard on all models)         
Pump-down Time: 255 L/sec turbomolecular pump: <3 minutes for air/water check, <90 minutes for quantitative stability         
Temperature Settings: SMARTsource: Settable 50 °C – 350 °C, GC transfer line: Settable 50 °C – 350 °C         
Carrier Gas Flow Rate: 255 L/sec turbomolecular pumping systems: 5 ml/min, 75 L/sec turbomolecular pumping systems: 1.5 ml/min         
Mass Calibrant: PFTBA (FC-43) gas, triazine (high mass applications), user selectable         
MS Isolation: SwaferTM Micro-Channel Flow Technology         
Field Upgrades: Positive/negative chemical ionization (255 L/sec), standard to large turbomolecular pump (75 L/sec to 255 L/sec), water cooling (all models)         
         
[u]Performance[/u]         
MS Data Collection: Full Scan, Selected Ion Monitoring (SIM), Simultaneous Full-Scan Selected Ion Monitoring (SIFI)         
Scan Rate: Fully variable up to 12,500 amu/sec         
Acquisition Rate: 100 points/sec (SIM)         
Functions/Run: 32 functions or 32 ions per function         
Linear Dynamic Range: 106 dependant on acquisition rate         
         
[u]Data system[/u]         
Software: TurboMassTM 6.0 designed for Windows 7 professional operating system         
Methods: Electronically transferable between all TurboMass and Clarus GC/MS systems         
GC Acquisition: Full control and data processing of a single GC/MS with up to 2 GC detectors         
MS Tuning: UltraTuneTM: Standard Tune environmental tuning optimized for EPA methods; Custom Tune user definable optimization         
Reporting: Included in all models with specialized reports for: Environmental and forensic applications.         
Optional Software:         
MS Libraries: NIST Mass Spectral Library, including AMDIS Deconvolution, Wiley Mass Spectral Library Maurer/Pfleger/Weber Library of Drugs,         
Pollutants, Pesticides, and Metabolites         
Advanced Processing: Ion Signature Deconvolution software         
         
[u]Physical[/u]         
Power: 120 VAC +– 10% @ 50/60 Hz +– 1% 1,000 VA, 230 VAC +– 10% @ 50/60 Hz +– 1% 1,000 VA         
Temperature: 10 °C – 30 °C (10 °C – 35 °C with water cooling option)         
Relative Humidity: 20 – 80 % non-condensing         
Weight: T and C models: 49.9 kg (110 lbs), S model: 46.7 kg (103 lbs), Forepump: 25.9 kg (57 lbs)         
Dimensions (HxWxD): Standalone MS: 50 x 32 x 77 cm (20 x 13 x 30 inches), Clarus SQ 8 T GC/MS (with Autosampler): 83 x 98 x 82 cm (33 x 39 x 32 inches)

NewsView All Articles

Looking for New or Used Equipment?

Shop on LabX

Shop on Labx.com