Biocode Hycel - Xenia

Manufactured by  Biocode Hycel
Follow this Equipment

True quantitative flow cytometer for haematology

High speed sequential measurements of impedance and light intensity at 12 m/s 
The two signals are match in less than 40 µs compare to 170 µs for other manufacturer.
Active Questions & AnswersAsk a Question

Recent Questions & Answers

avatar placemark

Updated byLabWrenchSiteAdmin

Hematology Analyzers Service ProvidersView All (1)

Request Support
Documents & Manuals

There are no Documents or Manuals available.

Features of Xenia
  • 5 part diff 

  • 27 Parameters 

  • Approx 60 tests/hour 

  • Programmable automatic rerun 

  • Extended linearity (patented) 

  • Automatic selective bidirectional transfer to the host 

  • No daily maintenance is required 

    Compressor 

  • Regulated high pressures 

  • Vacuum is constantly monitored and used only in correct conditions: other systems using

  • stepper motor can loose steps, i.e vacuum, without notice 

  • No risk of blockage with high pressure in front of the aperture (1.2 bar). 

    Metrology 

  • Absolute volume of the sampling loops onto the Ceramic Shear valve measured at workshop. 

  • High precision syringes 

  • Mechanical mixing of the sample provides excellent homogeneization. 

  • Constant dilution rate are warranted ±2% at 1/250 and 1/6000 for the whole production of instruments

General Specifications
Power Requirements 230 V/50 Hz or 115 V/60 Hz
Peak Power1000 VA
Additional Specifications

Throughput: 
about 60 Tests/hour 
 
Differentiation of WBC: 
Lym #, Mon #, Neu #, Bas #, Eos # 
Lym %, Mon %, Neu %, Bas %, Eos % 
Lx, Ly, Nx, Ny 
 
Measurement: 
WBC: Impedance and Laser scattering with double hydrodynamic focusing 
Hb: Photometry at 560 nm 
RBC, Hct, MCV, MCH, MCHC, RDV-CV, RDV-SD: Impedance with double hydrodynamic focusing 
Plt, Pct, MPV, PDW, by impedance with double hydrodynamic focusing 
Scattergram: electrical and optical measurement cell by cell 
Matching of the two measurements within 40µs only 
5 Histograms RBC, Plt, WBC/DC, WBC/LMN, WBC/NE) 
 
SampleSample Types: 
Patient, Plt only, Hb only, pre dilution with capillary, diluent, control, calibrator 
Final rates of the dilutions are identical for all modes, capillary mode included so that the same order of precision is matched whatever the mode used. 
 
Sample volumes: 
Capillary: 25 µL 
Direct (open or closed tubes): about 150 µL 
Manuel (open tubes only) about 180 µL 
Auto Sampler (closed only): about 280 µL (depending on the viscosity of the sample) 
 
Dimensions & weight: 
Analytical module : H 54 x L56 x P56 cm - 40 Kg 
Compressor : H17 x L42 x P60 cm - 18 Kg 
 
Options: 
Bar code reader (built in) 
Auto Sampler with or without built in bar code reader 
Bar code standards: Codabar, Code 39, Code 128, MSI, Interleaved 2 of 5