GE Healthcare - Typhoon™ FLA 9500
The flexible and versatile scanner for all imaging applications
The system is flexible and can be customized with lasers, filters and detectors for multiplex near infrared fluorescence with IRDye™.
Typhoon FLA 9500 provides 457/532/635 nm excitation and up to 10 µm resolution for gel, blot, tissue and array imaging. The system can be customized with 685/785 nm excitation, customized filters and detectors for multiplex and near infrared fluorescence applications.
Active Questions & AnswersAsk a Question
Regarding the price
Recent Questions & Answers
Updated byLabWrenchEditor
Need Equipment Support?
Documents & ManualsView All Documents
Features of Typhoon™ FLA 9500
Versatility: imaging of multifluorescent-,chemifluorescent-, radioisotope-labeled, and colorimetric samples
High resolution and quantitation: a pixel resolution of up to 10 Μm and a linear signal response over five orders of magnitude provides precise quantitation in gels, blots, tissue sections, and arrays
High sample throughput: scanning area of 40 × 46 cm enables simultaneous imaging of up to 20 gels or blots, measuring 10 × 8 cm in size. This facilitates comparisons among blots, and reduces workload and waiting time
Flexibility: optimized performance for new applications by adapting the system with stages, detectors, filters, and lasers
2-D DIGE imaging: simultaneously image two 2-D DIGE gels for differential expression studies
Visible and infrared fluorescence imaging: optional near infrared excitation for imaging IRDye™ and other infrared dyes
General Specifications
Depth | 800 mm |
Height | 400 mm |
Width | 900 mm |
Power Requirements | 100 - 240 VAC ± 10% |
Peak Power | 0.3 kVA |
Weight | 97 kg |
Additional Specifications
Detection modes:
Fluorescence, phosphorimaging, digitization, and chemiluminescence
Excitation wavelengths:
473 nm (blue LD laser),
532 nm (green SHG laser),
635 nm (red LD laser),
685 nm (optional red LD laser), and
785 nm (optional near IR LD laser)
Radioisotopes:
3H, 11C, 14C, 125I, 18F, 32P, 33P, 35S, 99mTc, and other sources of ionizing radiation
Dynamic range:
5 orders of magnitude
Bit depth:
16-bit
Scanning area:
40 x 46 cm
Pixel sizes:
10, 25, 50, 100, 200 µm and prescan 1000 µm
Standard filters:
IP (Phosphorimaging),
LPB (510LP), LPG (575LP), LPR (665LP),
LPR-Ch2 (665LP), BPB1 (530DF20),
and BPG1 (570DF20)
Optional filters:
BPFR700 (R715), BPFR800 (R810),
DBR1 (530DF20/665LP),
and DGR1 (570DF20/665LP)
Line frequency:
50/60 Hz
Temperature:
15°C to 30°C
Humidity:
20% to 70% (no condensation)