TA Instruments - Discovery TGA
The Discovery TGA features our industry-leading thermobalance, innovative IR-heated furnace and an autosampler...
The result is the ultimate in sensitivity, accuracy, resolution and temperature control... The Discovery TGA.
The Discovery TGA has replaced the Q5000 IR
Active Questions & AnswersAsk a Question
Recent Questions & Answers
Asked bytchowdh3
Need Equipment Support?
Documents & Manuals
There are no Documents or Manuals available.
Features
There are no Features available.
General Specifications
| Temperature Range | 28 to 1200 °C |
Additional Specifications
Temperature Range: Ambient - 1200°C
Isothermal Temperature Accuracy: ±1°C
Temperature Precision: ±1°C
Heating Rate Range:
0.1-500°C/min (linear)
>2000°C/min (ballistic)
Furnace Cooling (Forced Air/N2): 1200 to 35°C in >10 min.
Sample Weight Capacity: 750 mg
Dynamic Weighing Range: ±100mg
Weighing Accuracy: ±0.1%
Weighing Precision: ±0.01%
Signal Resolution: 0.001µ
Sensitivity: <0.1µg
Short-term Noise(rms): <0.03µg
Baseline Linearity (50-1000°C): <1µg
Baseline Drift(50-1000°C, 20°C/min), N2 purge, no baselines subtraction: <10µg