PID Analyzers - Model 312

Manufactured by  PID Analyzers
Follow this Equipment

Model 312 Portable GC

  • Features flame ionization and flame photometric detectors with flameout sensors that shut off the hydrogen if the flame goes out
  • Includes a 7-inch colour touch screen
  • Low and high set-points (depends on the options) can be custom programmed for each point and/pr compound.
Active Questions & AnswersAsk a Question

There are no current Discussions

GC Systems Service ProvidersView All (42)

Request Support
Documents & Manuals

There are no Documents or Manuals available.

Features of Model 312
  • PID Lamps: Our 9.5 and 10.6 eV lamps remain the best in the business since we developed them over 30 years ago.
  • FID: Our very sensitive FID has flameout sensor that will shut off the hydrogen if the flame goes out
  • FPD: Our very sensitive FPD has flameout sensor that will shut off the hydrogen if the flame goes out
  • Display: Our 7'' Color VGA display (640 x 480 resolution) is easy to read. The display is a touch screen
  • PC: Embedded Pentium Computer with > 1 GByte of RAM; data storage on 1 Gbyte USB memory stick
  • Operating system: Windows XP® embedded stored on 2.5 GByte smart card (the PeakWorks program and all data are on the memory stick
  • Peak Detection & System Operation: Controlled by our 32-bit PeakWorks software program with more than 20 years of development
  • ADC: Our 16 bit ADCs provide a resolution of 1 part in 64,000 and our signal algorithm minimizes noise by signal averaging in the msec range; the ADC can acquire positive or negative signals
  • Autoranging: The 312 has 3 ranges with computer controlled autoranging- x1, x 10, x 100
  • Autozero: Beginning of each run
  • Programmable Voltage/current output: 0-1VDC
  • Digital outputs: USB connection
  • Battery life: 3-4 hours; can operate off cigarette lighter adapter
  • Carrier gas: 50 liters at 1500 PSI; typical capillary column 15-20 cc/min; typical packed column 30-40 cc/min
General Specifications
DetectorPhotoionization detector (PID), Flame ionization detector (FID), Thermal conductivity detector (TCD), Flame photometric detector (FPD), Far UV absorbance (FUV) detector, Catalytic combustion detector (CCD)
Power Requirements 115 VAC, 0.5 amps (12 VDC 5 amps)
GC System TypeMicro
Weight27.5 lbs Fiberglass case
Additional Specifications

Detectors:       
Photoionization detector (PID), Flame ionization detector (FID), Thermal conductivity detector (TCD), Flame photometric detector (FPD), Far UV absorbance (FUV) detector, Catalytic combustion detector (CCD)        
       
Species measured:       
PID & FID: Organics (VOCs by PID or FID) and inorganic species (H2S, NH3, AsH3… ) that can be ionized by the PID lamp (9.5, 10.6 or 11.7 eV)        
TCD: Universal detector- fixed gases (O2, N2, CO…), organic and inorganic compounds, very sensitive for H2, He, Xe 
FPD: Selective response for S or P compounds       
Far UV: Nearly universal response for organic and inorganic compounds       
CCD: Combustible compounds such as hydrocarbons        
       
Detection Limit:       
PID- 0.5 ppb benzene       
FID- 50 ppb benzene       
TCD- H2- 10 ppm; O2 150 ppm       
Linear Dynamic Range:       
PID- 108       
FID- 2 x 106       
TCD- 2 x 104        
FPD- S- 103; P 105       
Far UV- 106       
CCD- 5 x 104       
Sensor(s):       
1 or 2 detectors       
Precision:       
+/- 1-2% of reading       
Response time:        
Depends on retention times       
Sample flow rate:       
> 0.5 LPM        
Injection:       
On-column syringe injection or automatic with 6 port gas valve (optional)        
Span Drift:       
<1% over 24 hours       
Zero Drift:       
NA- autozero on each run       
Dimensions:       
enclosure - 18.75''x 13.75'' x +7.5''        
Instrument operating conditions:       
5-40 degrees C, 0-95% RH (non-condensing)