Techne - Quantica
Quantica® has been designed with the advantage of having an open architecture and chemistry format.
Quansoft® software and reporting
Accompanying Quantica® is our easy-to-use, flexible software Quansoft®, which gives you the freedom to see your raw data, choose your disposables and choose your own chemistries!
- Ability to create, modify and save files using Quansoft® while a program is running on the unit
- Ability to run without initially defining your plate layout and analysis criteria - just pop a prepared plate in & run a program
- Ability to export unprocessed, raw data
- Install Quansoft® on multiple PCs without paying a license fee (only a simple registration is required)
- Ability to change from first order derivative to fit points in results analysis
Heated Lid
The flexible optical heated lid system is designed to minimize loss of sample and prevent sample condensation. Temperature controlled, it allows further method optimization where pre-incubation stages may be required (e.g. in reverse transcription techniques) or where changes in the heated lid characteristics are needed during the different stages of a program run.
Active Questions & AnswersAsk a Question
Recent Questions & Answers
Asked byMMartin
PCR / Thermal Cyclers Service ProvidersView All (14)
Documents & ManualsView All Documents
Features of Quantica
- Solid state white light source and PMT (photomultiplier tube) detector gives an impressive excitation range of 470nm to 650nm and a detection range of 500nm to 710nm.
- Multiplex - multiple wavelengths are detectable per sample using up to four paired excitation and emission filters housed in individual cartridge systems.
- Wide dynamic range - up to at least 9 orders of magnitude from starting copy number and high sensitivity, detecting down to 1nM fluorescein and single copy templates, depending upon the assay.
General Specifications
| Depth | 21 in |
| Height | 20 in |
| Real-Time PCR | 1 |
| Width | 16 in |
| Power Requirements | 100-230V, 5060Hz |
| Fluor. Excitation Range | 470 to 650 nm |
| Peak Power | 715 W |
| Weight | 55 lbs |
| Temp. Differential Range | 4 to 98 °C |
Additional Specifications
Block format: 96x0.2ml
Sample volume: 15 to 50ul
Heated lid: Adjustable between 100 and 115°C in 1°C increments, or off
Plate format: Low profile 96-well plate (skirted or non-skirted)