A & E Lab - AE-UV1802
Manufactured by A & E Lab
UV Vis Spectrophotometer AE-UV1802/AE-UV1802S
- With embedded computer and optimized optical system,convenient and easy to use.
- 320×240 dots array large screen graphic LCD display with abundant instructions,and curve and table directly displayed on screen.
- Basic photometric mode, Quantitative analysis. Qualitative analysis, Kinetics test, DNA/Protein Test and Multi-wavelength are available.
- Alternative fitting and correction methods for Quantitative Analysis.
- Re-scaling axes, curve smoothing,combination,zooming and overlap, 1st to 4st derivative for Qualitative analysis.
- Up to 1000 test data can be saved in local memory.
- Data output by optional built-in thermal printer or USB2.0 printer driver and PC software available.
Active Questions & AnswersAsk a Question
There are no current Discussions
UV/VIS Spectrophotometers Service ProvidersView All (32)
Documents & Manuals
There are no Documents or Manuals available.
Features
There are no Features available.
General Specifications
There are no General Specifications available.
Series Modals
A & E Lab - AE-UV1802
[+] View Model SpecificationsPhotometric Range | 0.3 to 3A ; 0-200%T; -9999 to 9999C; |
Wavelength Range | 190 to 1100 nm |
Photometric Noise | ±0.001A (500nm, 0A, 2nm SBW) |
Stray Light | = 0.05%T (220 nm, NaI; 340 nm, NaNO2) |
Wavelength Resolution | 0.1 nm |
Additional Model Specifications | Spectral Bandwidth: 1 nm or 2 nm (Optional) Wavelength Accuracy: ±0.1 nm(656.1 nm D2), ±0.3 nm(Whole wavelength range); Wavelength Repeatability: ±0.2 nm Photometric Range:0.3 to 3A ; 0-200%T; -9999 to 9999C; Photometric Accuracy: ±0.002A@0.5A , ±0.3%(0~100%T) Photometric Repeatability: ±0.001A@0.5 A, ±0.15%T(0~100%T) Baseline Flatness: ±0.001 A Stability: ±0.0005A/hr @500nm, 0A |
A & E Lab - AE-UV1802S
[+] View Model SpecificationsPhotometric Range | 0.3 to 3A ; 0-200%T; -9999 to 9999C; |
Wavelength Range | 190 to 1100 nm |
Photometric Noise | ±0.001A (500nm, 0A, 2nm SBW) |
Stray Light | = 0.05%T (220 nm, NaI; 340 nm, NaNO2) |
Wavelength Resolution | 0.1 nm |
Additional Model Specifications | Spectral Bandwidth: 0.5 nm, 1 nm, 2 nm, 4 nm, 5 nm Variable Wavelength Accuracy: ±0.1 nm(656.1 nm D2), ±0.3 nm(Whole wavelength range); Wavelength Repeatability: ±0.2 nm Photometric Range: 0.3 to 3A ; 0-200%T; -9999 to 9999C; Photometric Accuracy: ±0.002A@0.5A , ±0.3%(0~100%T) Photometric Repeatability: ±0.001A@0.5 A, ±0.15%T(0~100%T) Baseline Flatness: ±0.001 A Stability: ±0.0005A/hr @500nm, 0A |