Designed to handle the demands of smaller injection volumes—delivers microliter volumes with precision and accuracy.

Designed to handle the demands of smaller injection volumes—delivers microliter volumes with precision and accuracy. Injects 1µl (or less) from 96-well microplates, vials, or tubes.
Active Questions & AnswersAsk a Question

There are no current Discussions

Autosamplers Service ProvidersView All (1)

Request Support
Documents & Manuals

There are no Documents or Manuals available.

Features of Micro 215 Liquid Handler/Injec
      
  • Accommodates up to 17 standard or deep-well microplates   
  • Increases throughput with valve switching speeds five times faster than the standard 215 Liquid Handler/Injector   
  • Optional Peltier Racks for cooling and heating of samples in standard microplates   
  • Ideal for injecting 1 µL (or less) directly into the 841 Injection Module to minimize carryover and maximize precision; can also accommodate large volume injections without reconfiguring the instrument   
  • Add a second 841 Injection Module to increase throughput while preserving bench space  
General Specifications
Depth 24 in
Height22 in
Width 38.5 in
Power Requirements 100–120 V, 60 Hz / 200–240 V, 50 Hz
Weight39.9 kg
Additional Specifications

Manufacturing Standards  
Meets applicable safety and EMC certification standards; UL and CE certified  
Arm Speed X: >45.7 cm/sec (>18 in./sec)  
Y: >40.6 cm/sec (>16 in./sec)  
Communication Interface RS-232 or GSIOC; Four inputs (contact closure, TTL, or open-collector), four relay outputs, and one switched +24V DC 1A output  
Liquid Handling Reproducibility:  
1.37% CV with 50 µL of water  
Accuracy:  
99.7% with 50 µL of water  
Front Panel 8-character display, START and emergency STOP soft keys  
Horizontal Motion Strength X: 5 kg (11.1 lbs.)  
Y: 7 kg (15.6 lbs.)  
Injection Carryover† <0.05% with 100-µL syringe and total loop filling method  
Injection Reproducibility† 1.5% CV  
Injection Valve Switching Speed <100 ms  
Injection Volume 500 nL–5 mL  
Locator Plate Capacity Up to five Code 200-Series racks, up to seven Code 20- or 30-Series racks (or a combination of up to five racks of both types), or one Code 500-Series and one Code 200-Series rack  
Maximum Syringe Flow Rates for Water Syringe Size  
(µL) Max. Flow Rate  
(mL/min)  
100 3.37  
250 8.43  
500 16.87  
1000 33.75  
5000 100.00  
10,000 100.00  
25,000 100.00  
Probe Positioning Performance Accuracy: ±0.5 mm in X/Y dimensions, ±1 mm in Z dimension  
Repeatability: ±0.25 mm in X/Y/Z dimensions  
Probe Rinse Through a dedicated rinse station for rinsing the inside and outside of the probe; selectable rinse volume and flow rate. Optional inserts for level-sensing, nonleveling-sensing and flow rinse  
Sampler Type X/Y/Z with stationary rack design  
Software Via Gilson UniPoint™ System Software or TRILUTION™ Software  
Syringe Pump Integral, high-precision, single-piston  
Syringe Capacity 100, 250 or 500 µL; 1 or 5 mL  
System Resolution 0.1 µL  
Vertical Punch Strength 4.1 kg (11.0 lbs.)  
  
† Contact Gilson Customer Service to learn what methods and conditions were used to obtain the values.