METTLER TOLEDO - InPro 5000i

Manufactured by  METTLER TOLEDO
Follow this Equipment

Dissolved Carbon Dioxide Sensor

METTLER TOLEDO's InPro5000i dissolved carbon dioxide sensor is fully sterilizable in place (SIP) or in an autoclave. Optimum yield in bacteria fermentation is strongly related to the proper control of CO2 partial pressure. In cell culture applications, mammalian cells depend on a set of specific environmental conditions within a narrow definition in order to survive and to ensure optimal growth. INGOLD's extensive experience in offering high quality measuring solutions for biotechnology and pharmaceutical applications has been implemented in the new system.
Active Questions & AnswersAsk a Question

There are no current Discussions

Gas and Air Analyzers Service ProvidersView All (10)

Request Support
Features of InPro 5000i
On-site service with modular design  The interior body, a high performance pH electrode, can easily be replaced at the customer site. No need of decontamination and send to a METTLER TOLEDO facility for service.    High selectivity for CO2  No cross interference of volatile organic acids. A new sophisticated membrane allows only CO2 to pass through which guarantee accurate results.    MaxCertTM, a pratical validation standard  METTLER TOLEDO provides 3.1B, EHEDG and the production end control certificate free of charge to satisfy hygienic requirements and to guarantee traceability.
General Specifications

There are no General Specifications available.

Additional Specifications

Detection limit:  
lower detection limit: 10 mbar 
Accuracy:  
greater than +/- 10 %, 10...900 mbar; smaller than +/- 10 %, 900...1000 mbar 
Response time at 25 °C (air --> N2):  
90 % of final value smaller than 120 sec. (at 25 °C from air to CO2) 
Temperature range:  
0...60 °C (32...140 °F) for measurement; 130 °C (266 °F) for sterilization, in-situ or in an autoclave 
Pressure range (bar/psi):  
0.2...2 bar (3...29 psi) absolute for measurement 
Mechanical pressure resistance:  
3 bar (43 psi) at 25 °C (77 °F) 
Interference:  
known interferences: SO2, NH3, H2S 
Material in contact with medium:  
stainless steel DIN 1.4435 (similar AISI 316L), surface roughness N 5 Ra < 0.4Μm (16 Μinch) 
Material:  
membrane material: silicone reinforced, PTFE 
O-ring material:  
Viton®, silicone FDA compliant 
Temperature sensor:  
Pt1000 
Connector:  
AK9, Pg 13.5 thread 
O-ring material:  
Viton®, silicone FDA compliant 
Pressure range (bar/psi):  
0.2...2 bar (3...29 psi) absolute for measurement 
Temperature range:  
0...60 °C (32...140 °F) for measurement; 130 °C (266 °F) for sterilization, in-situ or in an autoclave 
Material:  
membrane material: silicone reinforced, PTFE

NewsView All Articles