Leica Biosystems - CV5030
Fully Automated Glass Coverslipper
Its capability of handling a large variety of slide racks from different suppliers makes the Leica CV5030 very flexible. All common mounting media, including xylene-free varieties, can be used.
The operator can choose wet or dry coverslipping. The CV5030’s performance achieves three goals: Highly consistent and reliable coverslipping quality, full adaptation to individual laboratory set-ups, integration into staining/coverslipping workstations to form a fully automated walk-away system.
Active Questions & AnswersAsk a Question
CV5030 Service Software & Manual
Recent Questions & Answers
Asked byJStone
Histology Equipment Service ProvidersView All (6)
Documents & ManualsView All Documents
Features of CV5030
- The dispenser parking station stores the nozzle in reagent – safely maintains mountant dispensing integrity.
- The easy-to-use control panel enables settings to be customized to meet your specific laboratory requirements – simply select the ideal set-up for different samples.
- The coverslip station accepts different sizes of coverglass, automatically discarding broken ones – Samples are safely and permanently covered, assuring consistent, high quality coverslipping.
General Specifications
Depth | 23.62 in |
Height | 21.65 in (with lid closed) |
Width | 14.57 in |
Power Requirements | 100-240V AC ± 10%, 50 to 60Hz, 100VA |
Weight | 125.55 lbs |
Additional Specifications
Ambient conditions
Ambient temperatures: +15 °C to +35 °C
Storage temperature range: +5 °C to +50 °C
Operating elevation: up to a max. of 2000 m above sea level
Relative humidity: max. 80%, non-condensing
Storage humidity: < 80%, non-condensing
Electrical data
Nominal supply voltages: 100-240V AC ± 10%
Nominal frequency: 50 to 60 Hz
Nominal power: 100 VA
Main fuses: 2x Automatic Circuit Breaker T 5 A
IEC 1010 classification:
Protection class 1
Pollution degree 2
Overvoltage installation category II
International Protection Class: IP20
Connections: RS 232C - communication interface
Uninterruptible power supply (UPS):
The UPS should be designed for a capacity of at least
200VA for a period of 5 minutes.
Admission: CE, CSA C/US
Additional data
A-weighted noise level: = 70dB (A)
Heat dissipation: 100J/s
Dimensions and weights
Height:
980mm (38.58 Inch) (with lid opened)
Width (from right foot to left foot): 370mm (14.57 Inch)
Depth (from rear foot to front foot): 525mm (20.67 Inch)
Width (CV5030 with TS5025 Transferstation): 560mm (22.05 Inch)
Weight (including shipping crate): 104kg (229.07 lbs.)
Software interlock: ON/OFF