Tecan - Cavro XLP 6000 pump
The most versatile 60mm stroke OEM syringe pump
The pump has been designed for ease of integration and low maintenance. The new hardware and firmware automatically detects the communication interface (RS 232, RS 485, or CAN) and communication protocol (OEM, Data Terminal). The Cavro XLP 6000 Pump also allows the user to reconfigure the pump to run different valve types and at different baud rates. The two digital inputs and three outputs for TTL level signals can be used for synchronization with external devices. Multi-pump communication is possible through RS 485 or CAN interface, addressing up to 15 modules through a single communication bus.
The Cavro XLP 6000 Pump family of products is UL recognized (standard UL61010A-1) and bears the UL designation label.
Active Questions & AnswersAsk a Question
Recent Questions & Answers
Asked bytri
Lab Pumps Service ProvidersView All (3)
Documents & Manuals
There are no Documents or Manuals available.
Features of Cavro XLP 6000 pump
- More flexibility with wide range of syringe volumes and valve options
- Robust and reliable workhorse
- Easy to integrate
General Specifications
Depth | 4.76 in |
Height | 10 in |
Width | 2.56 in |
Weight | 4.7 lbs |
Additional Specifications
Plunger Drive
Principle:
Stepper motor driven lead screw with optical encoder for step loss detection
Travel:
60 mm
Syringes:
50 µL, 100 µL, 250 µL, 500 µL, 1.0 mL, 2.5 mL, 5.0 mL, 10 mL, 25 mL
Valve/Valve Drive
Wetted Materials:
Plastic - Plug: FEP
Body: PCTFE
Options:
120° 3-port, 90° 4-port, T-valve, Y block, 90° 3-port distribution
Turn Time:
< 250 ms between adjacent ports
Fittings:
1/4-28” or M6 tubing fitting, 1/4-28” syringe fitting
Firmware:
Programmable ramps, cutoff speed, backlash compensation, plunger speeds, delays and loops, change speed on the fly, terminate moves, delays, diagnostics, absolute and relative positions
Power Requirements
Supply Voltage:
24 VDC ± 10%
Current:
Peak: 1 A
Resolution:
Standard:
6,000 (standard) / 48,000 (microstepping) increments
Speed:
1.2 seconds - 160 minutes per stroke
Precision:
= 0.05% C.V. within run at full stroke (at 23°C, with 500 µL syringe and above using deionized water)
Accuracy:
= 0.5 % deviation from theoretical result at full stroke (at 23° C, with 500 µL syringe and above using deionized water)
Interface
Type:
RS-232, RS-485, or CAN BUS
Baud Rate:
RS232/485: 9,600 or 38,000
CAN: 100Kb, 125Kb, 250Kb, 500Kb, and 1Mb
Format:
Data Bits: 8
Parity: None
Stop Bit: 1
Half Duplex
Addressing:
Maximum of 15 pumps can be addressed individually
Communications:
Data Terminal and OEM protocol (serial); “Cavro CAN” (CAN)
Environmental (Operating Mechanism)
Operating Temperature:
15° - 40°C (59° - 104°F)
Humidity (Non-Condensing):
20-80% RH at 40°C (104°F)
Safety and Regulatory:
The XLP is a UL recognized component. Tecan Systems UL customer file is E164638.
Compliance:
Tecan Systems' Quality System is ISO 13485 Certified.