GE Healthcare - IN Cell Analyzer 2200
IN Cell Analyzer 2200 is a super-fast and sensitive wide-field cell imaging system, fine-tuned to the needs of the...
Building on the extremely successful IN Cell Analyzer 2000 system, IN Cell Analyzer 2200 is designed to bring efficiency and flexibility to your imaging, making both simple and complex high-throughput, high-content assays an everyday reality.
With a scientific-grade CMOS camera and bright solid-state light source as standard, IN Cell Analyzer 2200 reduces exposure times and maximizes speed while delivering high-quality images. Confocal-like images can also be obtained using the rapid image restoration options in the control software.
IN Cell Analyzer 2200 can image a two-color 96-well plate assay in less than 2.5 min. This speed is achieved through the highly optimized optical system and highly efficient control software, helping you spend your time where it matters most—on your research.
Active Questions & AnswersAsk a Question
Recent Questions & Answers
Asked byriccio
DNA Sequence Analyzers Service ProvidersView All (6)
Documents & Manuals
There are no Documents or Manuals available.
Features of IN Cell Analyzer 2200
- Perform anything from investigative microscopy to automated high-content screening.
- Image organelles, cells, tissues, and whole organisms.
- Carry out fixed end-point assays as well as extended live-cell studies.
Designed for Flexibility IN Cell Analyzer 2200 incorporates intuitive and easy-to-use tools that simplify the imaging process for different sample types and enable a wide range of applications, including compound screening, predictive toxicology, protein localization and trafficking, functional studies, and stem cell analysis.
IN Cell Analyzer 2200 can also be configured to your specific needs. Choose from a wide range of optional modules and accessories to build the instrument you need now, or upgrade as your needs evolve.
Transform Images into Knowledge Obtaining images from IN Cell Analyzer 2200 is only the first step on the road to discovery through high-content imaging. Easy-to-use, flexible analysis and visualization tools are key to unlocking the full potential that high-content analysis offers.
IN Cell Investigator is designed to provide a comprehensive package of image analysis tools for a broad range of imaging experiments, suitable for users with basic as well as advanced knowledge of cellular image analysis.
General Specifications
Depth | 654 mm |
Height | 605 mm |
Temperature Range | 5 to 42 °C |
Width | 1140 mm |
Power Requirements | 100 to 240V @ 50 to 60Hz |
Weight | 50 kg |
Additional Specifications
Camera:
sCMOS
Pixel size: 6.5 Μm x 6.5 Μm
Pixel Layout: 2560 x 2160
Pixel used for imaging: 2048 x 2048
Complete Packsize:
1 Piece
Computer:
Windows 7 Professional Workstation (64 bit)
Environmental Control Module [CO2]:
A sample chamber maintains control of CO2 concentration and relative humidity (to minimize evaporation).
Complete environmental control prolongs cell health and viability, ensuring more reliable results from live-cell imaging assays performed over extended periods (optional).
Filter Selection:
Excitation Filters: 8
Focus:
Hardware: Laser sensor
Software: Contrast based algorithm
Z Stage Resolution: 25nm
Illumination:
Solid State Illuminator(SSI): 7 excitation channels
Lifetime: >10,000h
Transmitted light source: LED
Includes:
Hardware autofocus
Software autofocus
10× 0.45 NA objective, 20× 0.45 NA objective with automated spherical aberration collar (ASAC) correction
Standard polychroics, filter set
2D deconvolution software
Computer, monitor, keyboard, and mouse
Magnification:
Polychroic Control: Automated 2 positions
Objective Control: Automated 4 positions
2x/0.1 Plan Apo
4x/0.2 Plan Apo
10x/0.45 Plan Apo
20x/0.45 ELWD Plan Fluor
20X0.75 Plan Apo
40x/0.60 ELWD Plan Fluor
40x/0.95 Plan Apo
60x/0.70 Plan Fluor
60x/0.95 Plan Apo
100x/0.90 Plan Fluor
Plate Compatability:
SBS standard footprint microplates including 6-, 12-, 24-, 48-, 96-,384- and 1536-well plates and 1" x 3" standard slides
Plate Positioning [XY Stage]:
Resolution: 500 nm
Temp. Min. [For Kinetic Assays]:
Ambient +5 °C
Temp. Max. [For Kinetic Assays]:
42 °C