NuAire - In-VitroCell ES NU-5800

Manufactured by  NuAire
Follow this Equipment

Direct Heat CO2 Incubator

The In-VitroCell ES NU-5800 Direct Heat Laboratory CO2 Incubator is designed to provide a controlled in-vitro environment by providing the desired functionality needed to encourage cell culture growth for research at or near body temperature.
Active Questions & AnswersAsk a Question

Recent Questions & Answers

avatar placemark

Asked bygaurisrivas9293

CO2 Incubator Service ProvidersView All (28)

Request Support
Features of In-VitroCell ES NU-5800
  • NuTouch Electronic Control System   
  • 100% Stainless Steel Coved Interior Chamber   
  • Dual Temperature Sensor Probes   
  • Infrared (IR) CO2 Sensor   
  • Dual Sterilziation Cycles (NU-5810, 5820, 5830, 5840)   
  • Humidity Control System (NU-5820, 5840)   
  • O2 Control System (NU-5830, 5840)   
  • Four (4) Stainless Steel Shelves   
  • Eight (8) Stainless Steel Shelf Guides   
  • Four (4) Wall Brackets   
  • Field Reversible Outter Door   
  • Remote Alarm Output Contacts   
  • 4 to 20 mA Analog Output RS-485 Communication USB Port   
  • CO2 Sample Port   
  • Adjustable Leg Levelers   
  • Access Port and Plug with Breather Holes   
  • One (1) Water Pan   
  • One (1) 6.5 ft. / 2 m Electrical Cord
General Specifications
CO2 Incubator TypeDirect Heat
Depth 27 in
Height40 in
Internal Depth 20 in
Internal Height 28.5 in
Width 26.5 in
Internal Width 21 in
Chamber Volume200 L
CO2 SensorInfrared (IR)
CO2 Range0.1 to 20 %
Weight235 lbs
Maximum Shelf Load20 lbs per shelf
Shelf Dimensions19 ¼ x 18 ¼ (WxD)
Incubator Orientation Floor Standing
Additional Specifications

Utility Connections    
Gas Connections:    
0.25" (6.3 mm) Tubing   
Gas Input Pressure:    
20 PSIG (1.4 BAR) Input Pressure Maximum. Two Stage Gas Regulator Required   
   
Electrical Requirements    
Startup Power:    
625 watts   
Running Power:    
250 watts   
Heat Rejected:    
14 BTU / minute   
   
Temperature Control System   
Sensor Type:    
Precision Integrated Circuit   
Range:    
5°C to 55°C (5°C aboveambient to 30°C max. ambient)   
Accuracy:    
±0.1°C   
Uniformity:    
± 0.3°C at 37°C   
Recovery:    
0.12°C/minute Average   
Display Resolution:    
0.1°C   
   
CO2 Control System   
Accuracy:    
± 0.1%   
Recovery:    
Up to 5% (-0.50% / +0.20%) in 5 minutes average.   
Display Resolution:    
± 0.1%   
   
Warranty    
2 Year Parts and Labor (US & CA)   
3 Year Parts (Intl)   
   
Quality Asssurance:    
UL, UL-C, CE, Energy Saver, RoHS, Constant Contamination Control, Critical Cleaning Components

Looking for New or Used Equipment?

Shop on LabX

Shop on Labx.com