Roche - cobas e 411
Bench top analyzer for heterogeneous immunoassays.
For immunological analysis of serum or plasma.
Active Questions & AnswersAsk a Question
Recent Questions & Answers
Asked bykyoung hei Han
Need Equipment Support?
Documents & ManualsView All Documents
Features of cobas e 411
General Specifications
| Power Requirements | 100 –120 VAC 50/60 Hz single phase or 200–240 VAC 50/60 Hz single phase |
| Peak Power | 800 VA |
Additional Specifications
System:
Fully automated, random access system for immunoassay analysis.
It is available as both, a disk system and a rack system.
System components:
Analytical module including Window XP embedded operated touch screen PC
Sample handling module: rack or disk operated
Sample throughput:
Up to 88 samples/hr (theoretical max)
Test throughput:
Up to 88 tests/hr (theoretical max)
Number of channels:
18 channels/reagent slots for up to 18 different assays
Programmable parameters:
Max 60 assays definable via 2D-barcode (programming by loading)
Sample types:
Serum, Plasma, Urine
[u]Sample input/output[/u]
Load/unload capacity:
30 samples (disk)
75 samples on 15 racks
Rack:
RD standard 5 position rack
Rack types:
Routine, STAT, Control, Calibrator
STAT handling:
Any unoccupied position on the sample disk,
dedicated STAT port on rack feeder
[u]Sample container types[/u]
Primary tubes:
5–10 ml; 16x100,16x75, 13x100, 13x75mm
Sample cup:
2,5 ml
Micro cup:
not allowed
Cups on tube:
Cup on top of a 16 x 75/100 mm tube
Sample volume:
10 to 50 Μl per test, depending on assay protocol
[u]Operating conditions[/u]
Ambient temperature:
18 to 32 °C (64.4 °F to 89.6 °F)
Ambient humidity:
20% to 80%
Noise Output:
60 dbA (stand-by mode)
63 dbA (avg. during operation)
[u]Physical dimensions[/u]
Width:
1200 to 1700 mm (disk/rack)
Depth:
730 to 950 mm (disk/rack)
Height:
560 mm (w/o PC unit)
Weight:
Approx. 170 kg (disk) and 210 kg (rack)