New Brunswick Scientific - Innova 3100 Water Bath Shakers
Ideal for culturing a wide range of cell types at temperatures up to 100°C, as well as for staining, destaining and hybridization protocols.
Larger-capacity than many water bath shakers, the Innova 3100 easily accommodates gel trays, tubes and flasks.
Active Questions & AnswersAsk a Question
There are no current Discussions
Water Baths Service ProvidersView All (15)
Documents & ManualsView All Documents
Features of Innova 3100 Water Bath Shakers
- Adjustable Level Control Automatically Replenishes Evaporated Water, maintaining desired level without monitoring or refilling. Water-in and drain lines are each fitted with control valves to simplify cleaning, filling and complete draining
- Temperature Range of 5°C Above Ambient to 100°C provides for a wide range of laboratory work, using molecular and cell biology protocols. (Elevated temperatures attained with aid of an optional gable cover and with a bath liquid with a high boiling point to minimize evaporation)
- Optional Cooling Coil extends the range of applications to near and below ambient conditions
- Temperature Precisely Controlled within ±0.1°C in 30 - 40°C range
- Speed Range, 25 - 400 rpm, ±1 rpm
- Orbit: 1/2" (1.3cm) diameter orbit
- Acceleration Circuit Builds Up Speed Gradually, avoiding sudden starts and unwanted splashing
- Deep Bath with Rounded Corners and Baffles Allows Elevated Speeds with minimal turbulence or splashing
- One-Piece Seamless Bath is crevice-free, easy to keep clean; rust-resistant stainless steel
- Heat Loss and Evaporation Minimized with use of optional Plexiglas or stainless-steel gable covers
- Optional Mobile Cart, or Space-Saving Dolly that Fits Under Benches and desks, are available for use where bench space is limited
General Specifications
Depth | 27.5 in |
Height | 14.5 in |
Internal Depth | 17.5 in |
Plate Depth | 16.5 in |
Temperature Range | 5 to 100 °C |
Bath Volume | 15 L |
Internal Height | 8 in |
Maximum Shaking Speed | 400 rpm |
Plate Width | 12 in |
Timer | Programmable shaking periods from 0.1 hr. to 99.9 hrs. Triggers audible alarm and energizes status light at end of period. Can be deactivated for continuous operation |
Width | 17 in |
Internal Width | 13.25 in |
Maximum Temperature | 100 °C |
Power Requirements | 120V 50/60 Hz; 220/230V 50/60 Hz; 240V 50/60 Hz; 100V 50/60 Hz, 1,100 VA |
Orbital Diameter | 0.5 in |
Weight | 105 lbs |
Controls | Digital |
Motor | Triple-eccentric counterbalanced drive in cast-iron housing with 9 permanently lubricated ball bearings. Solid-state brushless D.C. motor |
Shaking Speed | 25 to 400 rpm |
Shaker Type | Incubating Shaker |
Additional Specifications
Operating Conditions: 0-60°c, 90% RH, non-condensing
Heaters: Low watt-density, silicone rubber mat heater with high-temperature safety thermostat cut off
Alarms, audible and visible: Indicate when speed deviates more than 5 rpm or temperature more than 1°C from setpoint, and when timer operation has expired. Also signals low-water level condition. Audible alarm may be muted.
LED display: Indicates speed in increments of 1 rpm, temperature in 0.1°C, running time in 0.1 hrs., and alarm conditions. Character height: 9/16" (1.4 cm)
Recorder output: 0 - 5 volt output for tracking temperature and speed with external recorder or computer
Safety: Acceleration circuit prevents sudden starts/stops, minimizing both splashing & mechanical damage. Safety shuts off temperature when high-temperature limit is exceeded
Bath: Deep-drawn, stainless-steel, seamless bath with radius corners. Adjustable level control from 0.5" to 3" above shaker platform.Water-in and drain lines with shut-off valves and overflow line provided. (17.5" L x 13.25" W x 8" H [43 x 35 x 20 cm])
Automatic filling: Bath automatically refillable to a volume of 15 L; requires connection to water inlet line
Setpoint retention: All setpoints and operating status are retained by non-volatile memory. Automatic restartafter power is restored (indicated by flashing display)