New Brunswick Scientific - Classic C-Line C76
Ruggedly built controlled-temperature water bath shaker designed for continuous-duty orbital shaking, 24 hours a day, year-in and year-out.
Microprocessor Control
The C76 microprocessor controller compensates automatically for changes in workload and voltage to maintain setpoints and control tolerances with unusual precision. In the event of a power outage, the C76 retains all of its memory, restarting automatically after power is restored.
Active Questions & AnswersAsk a Question
Recent Questions & Answers
Water Baths Service ProvidersView All (15)
Documents & ManualsView All Documents
Features of Classic C-Line C76
- Microprocessors lock onto precise speed and temperature
- Temperature-controlled within ± 0.5º over a range from 7ºC above ambient to 80ºC
- Powerful eccentric drive with variable speed AC motor built for continuous use
- One-piece seamless bath of heavy gauge stainless steel with rounded corners for ease of cleaning
- Acceleration circuit prevents sudden starts and splashing
- Large capacity water bath accommodates flasks, tubes and dishes for culturing, staining, destaining and hybridization. Full and half-size accessory platforms are available
General Specifications
Depth | 26.9 in |
Height | 18.5 in |
Plate Depth | 16.5 in |
Temperature Range | 7 to 80 °C |
Maximum Shaking Speed | 300 rpm |
Plate Width | 12 in |
Timer | Programmable shaking periods from 0.1 hr to 99.9 hrs. Shuts off agitation at end of period. Can be deactivated for continuous operation. |
Width | 21.4 in |
Maximum Temperature | 80 °C |
Power Requirements | 100/120V, and 230/240V, 50/60 Hz. 960 Watts |
Orbital Diameter | 0.5 in |
Weight | 115 lbs |
Controls | Digital |
Motor | Variable speed AC motor. Eccentric counterbalanced drive with permanently lubricated ball bearings. |
Shaking Speed | 50 to 300 rpm |
Shipping Weight | 145 lbs |
Shaker Type | Incubating Shaker |
Additional Specifications
Indication: Digital LED display in 1 rpm increments
Temperature regulation: Pl microprocessor feedback control with platinum RTD sensor, employing pulse-width modulation of heater
Temperature indication: Digital LED Display in 0.1°C increments
Heaters: Low-watt density resistance-type heaters
Alarms: Visible warning signal indicates when speed deviates more than 5 rpm, temperature more than 1°C from setpoints, and when timer has expired
LED display: Indicates speed, running time, alarm conditions and temperature. Displays readout of internal clock. Character height 3/8” (0.95 cm)
Safety: Sensor shuts off heaters if high temperature limit is exceeded
Setpoint retention: Setpoints and operating status retained by non-volatile memory. Automatic restart after power is restored. (Indicated by flashing display)