Arizona Instrument - Jerome J605 Hydrogen Sulfide Analyzer
The J605 is housed in a light and ergonomically designed case, setting a new precedence for low-level, portable monitors.
Active Questions & AnswersAsk a Question
There are no current Discussions
Elemental Analyzers Service ProvidersView All (6)
Documents & ManualsView All Documents
Features of Jerome J605 Hydrogen Sulfide Analyzer
- No PC Software Required
- 20,000 Data Points
- On-Board Data Logging Stores Date, Time and Locations
- USB Interface for Data Transfer
- Battery Powered Sensor Regeneration
- 24 Hour Battery Life
- Detection H2S Levels as Low as 3 ppb
- SCADA Interface Capabilities via 4-20ma
Functional Test Module (FTM) Verifies Instrument Functions Correctly Between Recommended Annual Factory Calibrations
General Specifications
Height | 6.5 in |
Power Requirements | 12 |
Weight | 5.4 lbs |
Additional Specifications
Resolution:
20 ppt (0.02 ppb) (range dependent)
Detection Range:
3 ppb (0.003 ppm) - 10 ppm in three graduated ranges
Data Storage Capacity:
20,000 samples
Operating Environment:
0 - 40°C non-condensing, non-explosive
Accuracy:
5 ppb ±1 ppb (range 0)
50 ppb ±3 ppb (range 0)
0.5 ppm ±0.03 ppm (range 1)
5.0 ppm ±0.3 ppm (range 2)
Typical Response Time:
Dependent on Range and Mode
MANUAL SURVEY MODE:
12-27 seconds
AUTO SAMPLE MODE:
12-52 seconds
Flow Rate:
150 ml/min (0.15 liters/min)
Power Requirements:
12 VDC for the instrument (provided by the internal battery, AC power supply/charger,external battery pack or car accessory cable
100-240 VAC, 47-63Hz, 3.2 A for the AC power supply/charger
Internal Battery Pack:
Rechargeable nickel metal hydride (NiMH)
Dimensions:
11" L x 6" W x 6.5" H
Output:
Digital: USB serial data to PC, printer or USB flash drive
Analog: 4-20mA current loop (requires external power source) accurate to 0.3% of output
Warranty:
1 year, factory parts and labor