KDS Legato 110 Single Syringe Infuse/Withdraw Programmable

Single syringe infuse/ withdraw with program option for 2 programs. 50 steps each. Accomodates syringes 0.5ul to 60ml. User definable flow rates with selectable target volume or time values to control the total infusion volume.
Active Questions & AnswersAsk a Question

There are no current Discussions

Lab Pumps Service ProvidersView All (3)

Request Support
Documents & Manuals

There are no Documents or Manuals available.

Features of Legato 110
  • Holds one syringe from 0.5ul to 60ml   
  • High resolution color touch screen   
  • Unparalled ease of use   
  • Touch pad "lock" feature   
  • LED light on front panel   
  • Full metal chassis   
  • Built in syringe table   
  • Up to 30 lbs linear force   
  • Advanced microstepping techniques   
  • Built in RS485 interface   
  • USB port   
  • I/O & TTL interface   
  • Continuous mode of operation   
  • Spill dam   
  • Programmable
General Specifications
Depth 7.5 in
Height9 in
Width 5 in
Power Requirements 100-240 VAC: 50/60 Hz, 50 W, 0.5 A fuse
Weight5.9 lbs
Additional Specifications

Accuracy:    
± 0.5%   
Reproducibility    
± 0.05%   
Syringes (Min/Max)    
0.5 ul to 60ml   
Flow Rate:    

  • Minimum (0.5 ul syringe): 1.28 pl/min   
  • Maximum (10ml syringe): 25.99 ml/min   
  • Maximum (60ml syringe): 88.28 ml/min   
    Display    
    4.3 WQVGA TFT Color Display with Touchpad   
    Non-Volatile Memory    
    Stores all Settings   
    Connectors:    
  • RS485 - IEEE-1394 6 pos   
  • USB - Type B   
  • I/O & TTL - 15 pin D-Sub Connector   
    Linear Force (Max)    
    13.6 kg (30 lbs) @ 100% Force Selection   
    Drive Motor    
    0.9 Stepper Motor   
    Motor Drive Control    
    Microprocessor with 1/16 microstepping   
    Number of Microsteps per one rev.of Lead Screw    
    15,360   
    Step Revolution    
    0.069 um/step   
    Step Rate - Minimum    
    27.5 sec/ustep   
    Step Rate - Maximum    
    26 usec/ustep   
    Pusher Travel Rate - Minimum    
    0.15 um/min   
    Pusher Travel Rate - Maximum    
    159 mm/min   
    Operating Temperature    
    4 degree C to 40 degree C (40 degree F to 104 degree F)   
    Storage Temperature    
    -10 degree C to 70 degree C (14 degree F to 158 degree F)   
    Humidity    
    20% to 80% RH non condensing   
    Mode of Operation    
    Continuous   
    Classification    
    Class 1   
    Pollution Degree    
    1   
    Installation Category    
    11   
    Regulatory Certification    
    CE, UL, CSA, CB Scheme, EU RoHS