Features of Zetasizer APS
The Zetasizer APS enables you to:
- Investigate your protein over a large number of environmental conditions with minimal user intervention.
- Determine whether a protein is well-behaved in a buffer or whether aggregates are present.
- Assess purity to screen for suitable crystallization buffers.
- Measure aggregation temperature (melting point) to predict or compare stability.
- Uses industry standard 96 or 384 well plates.
- Measures size, size distribution, aggregation temperature.
- Use multiple measurement protocols on a single plate.
- Graphical plate scheduler to simplify set up.
- Color coded results graphic to aid data interpretation.
- Complete plate method stored for subsequent plates.
- Plate sampling technology ensures highest result accuracy.
- 2 - 90°C temperature range for measurement.
- 4 - 40°C temperature range for protein storage.
General Specifications
| Depth | 410 mm |
| Height | 420 mm |
| Width | 500 mm |
| Power Requirements | 100VA |
| Weight | 22.5 kg |
| Light Source | Laser 830nm, Max 60mW |
Additional Specifications
Measurement type 1 Particle size and molecular size
Measurement range: 0.3nm – 2.0 microns* (diameter)
Measurement principle: Dynamic Light Scattering
Minimum sample volume: 20µL
Accuracy: Better than +/-2% on NIST traceable latex standards
Precision/Repeatability: Better than +/-2% on NIST traceable latex standards
Sensitivity: 0.1mg/mL 15kDa protein (Lysozyme)
Measurement type 2 Aggregation temperature (melting point)
Measurement range: 2ºC - 90°C +/- 0.1.**
Measurement principle: Dynamic light scattering
Minimum sample volume: 20µL
Temperature control range: 2ºC - 90ºC +/- 0.1.**
Compatible plate types: 96, 384 well (SBS standard) †
Light source: Laser 830nm, max 60mW
Laser safety: Class 1
Operating environment:
- Temperature: 15°C – 30°C
- Humidity: 35% - 80% non-condensing
Recommended computer specification: Intel i5 processor PC, 4GByte RAM, 500GByte hard disk drive, 1280 x 800 screen resolution (widescreen), or 1280 x 960 (4:3), CD-ROM drive, USB port
Operating platforms: Windows XP, Windows 7, compatible with both 32bit and 64bit operating systems