Features of TH/TC-G/I-180

User-friendly design • Standard environmental test chambers • High quality 304 stainless steel exterior and interior • Casters for easy transport and installation • Convenient maintenance - available to check total operating time • Enough chamber capacity up to maximum 1000L • Select your optimum chamber from a full variation

  • The series provide three variations in temperature (& humidity) range of -40 to 150?, -40 to 120?, -60 to 150?/ upmost 20 to 98% RH, and five capacities of 180L, 300L, 408L, 800L, and 1000L, with a total 15 models altogether • Exclusive door lock & trim
  • Vice lock type door catch and foamed silicone packing ensure perfect chamber seal
  • Door open warning alarm • Inner door
  • Optional inner glass door has four hole to handle inside specimen without opening the door (Except for 800L and 1000L) • Ventilation
  • Ventilation vent is located at the upper right side of the unit • Recorder (optional)
  • 6-point dots type recorder with digital display • Cartridge type tank for easy water supply
  • Maintenance can be done easily from the front side
  • A center window allow to check the amount of water remaining easily • Viewing window for observation
  • A tempered window provides a clear view of your specimen during testing
  • Improved lighting provided • Access port for running in wires
  • Basic hole of 50mm diameter cable port is standard equipped on left side of chambers for wiring to the specimen. We also provide additional 50 or 80mm diameter cable port (Optional) • Easy cleaning of condenser air-filter
  • To prevent deterioration of refrigerator, keep filtering dust by periodical filter cleaning
General Specifications

There are no General Specifications available.

Additional Specifications

Chamber Volume (L/cu.ft.): 180/6.4
Control System: BDBC2 system - Balanced Dehumidifying & Bath Cooling Control
Controller: Programmable, Touch screen type controller (5.7" B/W monitor)
Temperature
Range (°C/°F):-35 to 150 / -31 to 302 - without humidity * Best programmable control range: -30 to 140 / -22 to 284
Flucuation (±°C/°F) at 55°C/60%RH: 0.1/0.18
Variation1)(±°C/°F) at 55°C / 60%RH: 0.6/0.18
Time
Heating-without humidity (Min.): 50 From -40°C to 120°C (-31°F to 248°F)
Cooling-without humidity (Min.): 60 From 20°C to - 35°C (68°F to -31°F) on single point control *
Humidity
Range (%RH):
40 to 85% RH at 15°C (59°F)
25 to 98% RH at 25°C to 80°C (77 to 176°F)
25 to 80% RH at 25 °C to 80°C (77 to 176°F)
Variation1) (±%RH) at 60%RH / 55°C: 3.0
Refrigeration
System: Mechanical single stage refrigerator system (air-cooled condenser)
Refrigerant: R-404A (CFC-free)
Capacity(W): 1500
Cooler: Plate fin cooler
Heater
Material: Incoloy sheath heating elemtent
Drying Bulb (W): 2000
Humidity Bulb (W): 1500
Sensor
Temperature: Pt. 100
Humidity: Electronic Sensor- free of drying a wet pack and good reaction time
Material
Interior: Stainless steel #304
Exterior: Stainless Steel #304
Insulation: Polyurethane foam 100mm (3.94")
Door: Glass Wool
AIr Flow: Tangent Blower (90W)
Water Tank
Capacity: 15L (0.53cu.ft)- Cartridge type
Water Quality: Ph6.2 to 7.2/ Electrical conductivity 20/cm to below
Dimensions
Interior (mm/inch): 500 x 630 x 600 / 19.7 x 24.8 x 23.6
Exterior(mm/inch): 1025 x 1035 x 1530 / 40.4 x 40.7 x 60.2
Viewing Window (mm/inch: 200 x 300 / 7.9 x 13.0
Cable Access Hole: Basicaly one hole provided/ Ø 50 or 80mm (2.0 or 3.2")
Electrical Requirements (230V): 60 Hz,/ 1ph/23.0A