PHCBi - MCO-80IC
CytoGROW Reach-IN Incubators
Active Questions & AnswersAsk a Question
There are no current Discussions
CO2 Incubator Service ProvidersView All (28)
Documents & ManualsView All Documents
Features of MCO-80IC
- Dual Infrared (IR2) Sensors for Optimal CO2 Control
- InCu-saFe® Copper Stainless Steel Alloy
- SafeCell UV Protection
- Direct Heat and Air Jacket System
- Integrated Microprocessor Control
General Specifications
| Depth | 33.6 in |
| Height | 80.3 in |
| Internal Depth | 27.3 in |
| Internal Height | 60 in |
| Width | 33.8 in |
| Internal Width | 31.7 in |
| Chamber Volume | 851 L |
| CO2 Range | 0 to 20 % |
| Weight | 275 kg |
Additional Specifications
OUTER DOOR
Double paned glass door with heater and outer door latch
SHELVES
30.6" x 25.9" (776 x 659 mm)
ACCESS PORT
[i]Inner Diameter:[/i] 40 mm (1.57 inch), 2 locations on right and left sides
INSULATION
Rigid-polyurethane foamed-in place
UV LAMP
Optional
TEMPERATURE CONTROL RANGE
Ambient temperature +º5C to 50ºC ([i]ambient temperature:[/i] 20ºC to 35ºC)
TEMPERATURE UNIFORMITY
±0.50ºC ([i]ambient temperature:[/i] 25ºC, [i]setting:[/i] 37ºC, [i]CO2:[/i] 5 %, no load)
MAXIMUM POWER CONSUMPTION
Max 1752 W
CO2 CONTROLLER
PID Control System
CO2 GAS PRESSURE
0.1 MPaG (inject):, 15 psiG)
CO2 VARIATION
±0.15% ([i]ambient temperature:[/i] 25ºC, [i]setting:[/i] 37ºC, [i]CO2:[/i] 5% no load)
CO2 SENSOR TYPE
Dual Infrared Sensor
DECONTAMINATION METHODS
24 hr full chamber UV decontamination (SafeCell Ultraviolet Light)
DISPLAY
Digital Touchpad
HUMIDIFYING SYSTEM
Sharing with interior (SUS304)
INTERIOR SURFACES
Stainless steel containing copper