Microtrac - BELSORP-MAX

Manufactured by  Microtrac
Follow this Equipment

Designed to measure a wide range adsorption isotherm for surface area and pore size distribution analysis.

Designed to measure a wide range adsorption isotherm for surface area and pore size distribution analysis.

The BELSORP–max utilizes the proprietary AFSM method for free space measurement, which produces a highly accurate measurement. The BELSORP-max can also measure vapor adsorption and perform chemisorption analysis.

Active Questions & AnswersAsk a Question

There are no current Discussions

Particle Size Analyzers Service ProvidersView All (2)

Request Support
Features of BELSORP-MAX
  • Specific surface area – 0.01 m²/g (N2), 0.0005 m²/g (Kr)
  • Pore size distribution
  • Proprietary AFSM measurement technique
  • Wide pressure range measurement – p/p0 10-8 ~ 0.997 (N2 at 77K, Ar at 87K)
  • Useful for Micropore analysis
  • Adsorption isotherm of a variety of gases(N2, AR, CO2, CH4, NH3) and vapors (H2O, Toluene, Benzene, Alcohol) can be measured
  • Simultaneous three sample measurement
  • Chemisorption option is available
  • TPD/TPR/TPO, pulse chemisorption measurement with TPD pro option
  • BELCryo option enables low-temperature measurements from 4.2k to 373k
General Specifications
Depth 580 mm
Height850 mm
Width 656 mm
Additional Specifications

Measurement principle: Volumetric adsorption method + AFSM™   
Adsorptive: N2, Ar, Kr, NH3, CO2, H2, Co, O2, CH4 and other non-corrosive gases   
H2O, MeOH, EtOH, C6H6 and other non-corrosive vapors   
Analysis Port:   
STD mode (p/p0=10 -4 -0.997) : 3 port   
High accuracy mode (p/p0=10 -4 -0.997) : 2 port   
High accuracy mode (p/p0=10 -8 -0.997) : 1 port (OP : 2 port)   
Specific surface area:   
0.01 m 2 /g and above (N2/77K, Ar/87K) 0.0005 m 2 /g and above (Kr/77K)   
Pore size distribution:   
0.35-500 nm in pore diameter   
Pressure transducer:   
133kPa (1000mmHg) 0.25% of F.S. x 5 units    
1.33 kPa (10mmHg) 0.5% of R. x 5 units (OP : 3 units)   
0.0133kPa(0.1mmHg)0.15% of R.x1unit(OP:2 units)   
Pressure transducer:   
1.6 x 10-6Pa (p/p0=1.6 x 10 -11 , N2 at 77K, Ar at 87K)-24bit A/D converter   
Pressure transducer:   
40oC   
Dewar Vessel:   
Volume:   
2.6L   
Holding time:   
60H