Varian SpectrAA 220G Graphite Furnace Atomic Absorption Spectrometer

Varian SpectrAA 220G Graphite Furnace Atomic Absorption Spectrometer. 
PC-controlled double beam AA spectrometer. The SpectrAA 110/220 optics feature 
four lamp positions with automated wavelength and slit selection, and high intensity deuterium background correction as standard. Lamps are mounted in fixed positions, and a rotating mirror is used to select the operating lamp, eliminating the need for a conventional turret. 
For maximum stability in an automated sequence, a software controlled next-lamp warm up facility is provided.
Active Questions & AnswersAsk a Question

Recent Questions & Answers

avatar placemark

Updated byybijandi

avatar placemark

Updated byLabWrenchSiteAdmin

Need Equipment Support?

Request Support
Features

There are no Features available.

General Specifications
Depth 58 cm
Height59 cm
Width 79 cm
Power Requirements 100
MonochromatorCzerny-Turner
Weight56 kg
Wavelength Range185 to 900 nm
Additional Specifications

MONOCHROMATOR Czerny-Turner with 1,200 lines/mm, holographic quartzovercoated grating, automatic wavelength & slit-size control   
Focal length, mm (DIMENSIONS (HXWXD), MAIN) 250   
WAVELENGTH RANGE, nm (UV RADIATION INTENSITY) 185-900   
RESOLUTION, nm 0.2, 0.5, 1; 0.5 reduced   
Scattering,nm/mm 3.2-2.4   
DETECTION LIMITS,ppb    
As (DETECTION LIMITS,ppb) 0.22   
Cu (DETECTION LIMITS,ppb) 0.15   
Ni (DETECTION LIMITS,ppb) 0.38   
Pb (DETECTION LIMITS,ppb) 0.1   
Se (DETECTION LIMITS,ppb) 0.23   
OTHER INFORMATION (STAINING POSTS) Windows-based software; multitasking; multielement; software choice of 11 QC tests or full EPA and userdefinable tests; optional software to assist with compliance with FDA 21 CFR Part 11; range of optional accessories, including sample introduction pump   
FORMAT (WAVELENGTH, nm) 1 channel, double beam - for more information please see comments.1 channel, double beam, time-shared, rotating-beam combiner   
BACKGROUND CORRECT (DETECTION LIMITS,ppb) Fast (2 msec) deuterium arc lamp to 2.3 absorbance units, 185-425 nm, electronic beam balance 1   
DISPLAYED INFORMATION (DETECTION LIMITS,ppb) Video displayemission, absorption, peaks, concentration, bracketing standards, standard additions, direct absorbance. With automatic attenuation   
Working curve (DETECTION LIMITS,ppb) Blank and 10 standards, 8 algorithms Additional readouts include corrected signals andcalibration graphs.   
Memory Unlimited methods, worksheets (includes labels, signal graphics), full export to LIS, spreadsheet 3   
Integration, sec (DETECTION LIMITS,ppb) 0.1-30 (1-20 replicates), PH, PA, PrOMT system, integrate repeat 4 Additional memory features include printer and OLE.   
Graphics (DETECTION LIMITS,ppb) Color CRT (SVGA) PrOMT is a feature that allows the user to specify the precision required; the instrument measures the sample for the time required to achieve thatprecision.   
PRINTER (EXTERNAL BEAM) Optional laser or ink jet, any printer supported by Windows   
Microcomputer (DETECTION LIMITS,ppb) Windows 98/ME/NT/ 2000/XP   
INTERFACE PROTOCOL (REPORT DISPLAY) PCI-IEEE standard   
DIMENSIONS (HXWXD) CM, (IN) (DISPLAY) 59 x 79 x 58 (23.2 x 31 x 23)   
WEIGHT, kg (lb) (DISPLAY) 56 (123)   
POWER NEEDED 100/120, 220/230/240   
Hz (OPERATING) 50/60   
HOLLOW-CATHODE LAMPS (DETECTION LIMITS,ppb) 4 lamps, automaticselect mirror, warmup capability, coded, noncoded; optional UltrAA (boosted) lamp