Gold Film Hydrogen Sulfide Analyzer
The portable Jerome® 631 Hydrogen Sulfide Analyzer displays low-level concentrations in just seconds. It offers an analysis range of 0.003-50 ppm for odor and corrosion control, safety, and leak detection in such industries as wastewater treatment, oil and gas, pulp and paper, and farming. This simple-to-use instrument utilizes an internal rechargeable battery pack or AC power, and is easily carried to suspected sources of hydrogen sulfide for detection and measurement. Locked in survey mode, the 631-X automatically displays hydrogen sulfide concentrations as quickly as every 3 seconds.
Active Questions & AnswersAsk a Question
There are no current Discussions
Elemental Analyzers Service ProvidersView All (6)
Documents & Manuals
There are no Documents or Manuals available.
Features of Jerome 631 Hydrogen Sulfide Analyzer
- Meets EPA & ATSDR Cleanup Levels
- Optional Data Logging
- Rugged design | Easy to operate
- Inherently stable gold film sensor
- Pressure sensitive membrane switch operation
- Rechargeable internal battery pack for portability
- Wide detection range allows for multiple applications
- Automatic LCD backlight during low light conditions
- Survey mode for rapid source detection of hydrogen sulfide concentrations
- Microprocessor ensures a linear response throughout the entire range of the sensor
- Functional Test Module (FTM) verifies instrument functions correctly between recommended annual factory calibrations
General Specifications
| Depth | 13 in |
| Height | 4 in |
| Width | 6 in |
| Power Requirements | 100-120 V~, 50/60 Hz |
| Weight | 8 lbs |
Additional Specifications
Resolution:
0.001 ppm
Detection Range:
3 ppb (0.003 ppm) - 50 ppm in four graduated ranges
Data Storage Capacity:
With optional data logger: 1,000 data points
Operating Environment:
0 - 40°C Non-Condensing, Non-Explosive
Accuracy:
0.05 ppm ±0.003 ppm (range 0)
0.5 ppm ±0.03 ppm (range 1)
5 ppm ±0.3 ppm (range 2)
25 ppm ±2 ppm (range 3)
Typical Response Time:
Dependent on Range and Mode
SURVEY MODE: 3-20 seconds
SAMPLE MODE: 13-30 seconds
Flow Rate:
150 ± 10 ml/min (0.15 ± liters/min)