PerkinElmer - Optima 3000
Manufactured by PerkinElmer
The Perkin-Elmer OPTIMA 3000 is based on a unique optical design optimized for use with an ICP. It uses an Echelle...
The Perkin-Elmer OPTIMA 3000 is based on a unique optical design optimized for use with an ICP. It uses an Echelle polychromator with a Segmented Array CCD. The detector consists of discreet subarrays which are strategically positioned to measure over 5000 emission wavelengths between 167 and 782 nm. The result is a simultaneous spectrometer that has unmatched resolution of 6 pm in the spectrally congested UV region.
Active Questions & AnswersAsk a Question
Recent Questions & Answers
ICP Systems Service ProvidersView All (15)
Documents & Manuals
There are no Documents or Manuals available.
Features
There are no Features available.
General Specifications
Depth | 76 cm |
Height | 155 cm |
Width | 127 cm |
Power Requirements | 00-240V, 30 A line, single phase, 50 or 60 Hz |
Additional Specifications
Depth with shelf - 103 cm (40.5 inches)
Weight: 454 kg (998 lb.)
Automation: Computer control of all spectrometer functions, RF generator and plasma gas flows over IEEEE-488 interface
Environmental: 15 to 35°C with <2.8 °C change per hour; relative humodity between 20 and 80% noncondensing