The DVR 2 is a fully electronic, versatile vacuum gauge for the measuring range from atmospheric pressure to 1 mbar. It...

The DVR 2 is a fully electronic, versatile vacuum gauge for the measuring range from atmospheric pressure to 1 mbar. It has an integrated ceramic diaphragm vacuum sensor with capacitive readout. The sensor measures independently of the type of gas and is exceptionally resistant to chemicals. It features outstanding measuring accuracy, very low temperature depencence and excellent long-term stability. A particular advantage of the DVR 2 gauge is its cordless, battery-powered operation. Its simple operation and the combined digital / analog vacuum reading for precise readout and quick trend detection, respectively, make the DVR 2 very versatile.
Active Questions & AnswersAsk a Question

There are no current Discussions

Vacuum Gauges / Controllers Service ProvidersView All (5)

Request Support
Features of DVR 2 Cordless Vacuum Gauge
  • High electromagnetic compatibility: Low emission level near detection limit, high degree of immunity to electromagnetic interference for use in industrial environments   * Large analog and digital vacuum display: Quick trend detection, precise readout   * User-selectable pressure units (mbar, hPa, Torr)   * Digital readjustment   * Capacitive alumina ceramic diaphragm vacuum sensor with excellent chemical resistance, accuracy and long term stability
General Specifications
Depth 115 mm
Height66 mm
Width 115 mm
Additional Specifications

Upper measuring limit mbar/torr    
1080 / 810   
Lower measuring limit mbar/torr    
1 / 1   
Measurement principle    
Ceramic diaphragm (alumina), capacitive, gas indep., absolute pressure   
Measurement uncertainty    
< +- 1 mbar/hPa/torr / +- 1 digit (after adjustment, constant temp.)   
Temperature coefficient    
< 0.07 mbar/hPa/0.05 torr /K   
Ambient temperature range (operation) °C    
10 - 40   
Ambient temperature range (storage) °C    
-10 - 60   
Max. media temp. for continuous operation / short times °C    
40 / 80   
Automatic switch-off    
User-selectable: 1-1000 min (default 15 min) or continuous operation   
Measurement cycle    
User-selectable: Automatic or 1 x per 3s, 1 x per 1s, 3 x per 1s   
Material of outer housing    
Robust plastic housing with good chemical resistance   
Degree of protection    
IP 40   
Power supply    
9 V Lithium battery, 1.2 Ah Ultralife U9VL   
Typical battery lifetime    
8.000 h, stand-by: 40.000 h