PerkinElmer - FIAS 100/400
Increase productivity on PerkinElmer AA, ICP-OES or ICP-MS systems
Sample handling is an important aspect of any successful trace elemental analysis. On-line procedures for sample preparation and delivery are increasingly becoming methods of choice as the demands for productivity increase. Flow injection atomic spectroscopy (FIAS) is a superb complement to PerkinElmer® atomic absorption (AA), inductively coupled plasma optical emission spectroscopy (ICP-OES) or inductively coupled plasma mass spectrometry (ICP-MS) systems as a means of sample preparation and delivery. A fully integrated and automated system, the FIAS can dramatically increase laboratory productivity and capability.
Active Questions & AnswersAsk a Question
Replicate measurements with FIAS 400
Recent Questions & Answers
Sample Preparation Service ProvidersView All (4)
Documents & ManualsView All Documents
Features of FIAS 100/400
- Automated hydride generation and amalgamation for improved detection limits for hydride-forming elements and mercury
- Analyte preconcentration or matrix removal for improved detection limits in difficult samples
- On-line addition of sample additives and internal standards for reduced sample-preparation time
- Sample pipelining to reduce washout times and improve stability for high-concentration samples
General Specifications
Depth | 41 cm |
Height | 18 cm |
Width | 42 cm |
Power Requirements | 115 VAC or 230 VAC |
Weight | 11.2 kg |
Additional Specifications
System description
Automated flow injection system for use with Flame AA, cold-vapor and hydride-generation AA, graphite furnace AA, ICP-OES and ICP-MS systems. The FIAS 100 system is provided with 1 pump and the FIAS 400 systems are provided with 2 pumps.
System control
Fully automated operation with computer-controlled spectrometer systems through control and samples windows integrated into PerkinElmer instrument-control software.
Transport of reagents
Stepper-motor driven peristaltic pump(s) with a maximum of 8 channels each for tubing with 0.13 to 3.18 mm inner diameter. Pump speed is selectable from 20 to 120 rpm.
Sampling and FI switching
Valve with 5 ports and exchangeable sample loops. An optional 8-port valve is flow switching available for the FIAS 400MS system. All systems have exchangeable sample loops of various volumes.
Gas supply
Carrier gas stream (argon) flow controlled from 40 to 250 mL/min by flow meter. Required inlet pressure: 320-400 kPa. Automatic switch-off after operating pauses of more than 10 min.
Absorption cell
Quartz cell is required for mercury/hydride AA analysis.
Cell heating
The cell can be heated using the air/acetylene flame or an optional electrically heated mantle(100 to 1000 °C).
Data connection
Built-in IEEE-488 and RS-232C interfaces. Addressing and switchover between IEEE and RS-232C by DIP switches.
Regulatory compliance
Classified as a laboratory instrument. Complies with the applicable European Union directives and standards for safety and electromagnetic compatibility for CE marking. The instrument has been developed and is produced in compliance with ISO 9001.