Baker - HypoxyCOOL
Accelerated Oxygen Conditioning for Tissue/Cell Culture Media
Active Questions & AnswersAsk a Question
There are no current Discussions
Tables / Workstations Service ProvidersView All (3)
Documents & Manuals
There are no Documents or Manuals available.
Features of HypoxyCOOL
- Improve research results – Avoid abnormal cell interactions and increase cell viability by simulating in vivo-like oxygen conditions directly in the growth media. * Enhance experimental control – Protocol-driven, tested solution helps eliminate media conditioning errors and preserves media integrity. * Increase productivity – Fast and efficient conditioning process reduces oxygen concentration to 2% within three hours. Conditioned oxygen level stays low for up to 21 days, when refrigerated.
General Specifications
Depth | 629 mm |
Height | 886 mm |
Width | 514 mm |
Weight | 91.6 kg |
Additional Specifications
Performance
Refrigeration Temp Range: 2°C - 6°C (35.6°F - 42.8°F)
Chamber O2 Setpoint Range: 0.5% - 15.0%
Run Time Set Point Range: 30 mins - 16 hrs
Shaker Speed Set Point Range: 70 RPM - 120 RPM
Gas Inlet Pressure: 30 psi - 70psi (2.0 bar - 4.8 bar)
Gas flow rate: 0.88 ft3/min - 1.24 ft3/min (25L/min - 35L/min)
Max Run Time: 16 hrs
Max Media Conditioning Capacity: 9 x 500ml bottles
Internal Dimensions - Refrigeration
Height: 349 mm (13.75")
Width: 410 mm (16.125")
Depth: 381 mm (15.0")
Electrical
Typical Motor Current: 4.5A (115V), 2.5A (230V)
Typical Operating Current: 6.6A (115V), 3.5A (230V)
Power Consumption: 520W (115V), 470W (230V)
Service Requirements (115V): 120V AC, 15A, 60 Hz, 8.2A minimum circuit ampacity
Service Requirements (230V): 230V AC, 10A, 50 Hz, 4.8A minimum circuit ampacity
Circuit Protection (115V): Maximum overcurrent protective device: 20A
Circuit Protection (230V): Maximum overcurrent protective device: 16A
Power Inlet: IEC 60320 type C-14 power inlet. Power cord ordered separately as required for locale.