Pharmacia PhastSystem high speed electrophoresis system The Pharmacia Phast System consists of a Separation-Control unit for system control and electrophoresis, and a Development Unit for gel staining. Electrophoresis is performed on a thermostatic plate capable of maintaining temperatures accurately from 0 to +70 °C. Multi-step programming capability allows gels to be pre run, loaded and run automatically. Controlled parameters include temperature, voltage, current, power, and volt-hours. The whole idea behind the Phast System is extreme ease of use and speedy runs. The stainless steel development unit is capable of processing two gels simultaneously at temperatures from ambient to +50 °C. Up to nine different reagents can be used in a single processing program.
Active Questions & AnswersAsk a Question
Recent Questions & Answers
Asked byDAL
Need Equipment Support?
Documents & Manuals
There are no Documents or Manuals available.
Features of Phast system
- Separation and control unit houses the separation bed and the programmable power supply.
- Separation bed can accommodate two gels simultaneously, offering a sample capacity of up to 24 samples,
- Cooled or heated accurately in the temperature range 4 to 30C.
- Extremely high field-strengths, up to 500V/cm on IEF gels and 120V/cm on SDS gels.
- Programming of up to 9 different separation protocols, each with up to 9 program steps.
- The stainless steel development chamber can process one or two gels at a time, at temperatures up to 50C.
- Rotating gel holder allowing maximum gel exposure and short development times (30-60 minutes).
- Voltage range: 10-2000V
- Current range: 0.1-50mA
- Power range: 0.1-7.0W
- Volthour integrator range: 1-9999 volthours/step
- Capacity: 1 or 2 gels.
- Volume 70ml solution
- Number of ports: 9, plus 1 for waste.
- Agitation: rotation
- Development: Solutions automatically pumped.
General Specifications
Depth | 60 cm separation, control and development units |
Height | 28 cm separation, control and development units |
Tray Capacity | 2 |
Width | 76 cm separation, control and development units |
Weight | 9 kg separation, control and development units |
Electropheresis Type | Gel |
Orientation (H/V) | Horizontal |
Additional Specifications
Separation and Control Unit
Capacity 1 or 2 gels
Keyboard 31 keys with audible tone
Display 40-digit alphanumeric liquid crystal
LEDs 4 LEDs for status information
Programmable parameters Voltage, current, power, temperature, duration in volt-hours, sample application and an extra alarm for 9 programs of 9 steps
Voltage range 10-2000 V
Current range 0.1-50.0 mA
Power range 0.1-7.0 W
Volt-hour integrator range 1-9999 Vh/step
Separation temperature range 0-70 C
Alarm Audible alarm at end of separation
Dimensions (WxDxH) 46 x 30 x 14cm
Weight 6.2 kg
Battery back-up Lithium battery
Safety features Short-circuit protection; protected against power disturbances and static discharges; lid power interlock and alarm
Development Unit
Capacity 1 of 2 gels
Volume 80 ml solution
Number of ports 9, plus 1 for waste
Agitation Rotation
Development Solutions automatically pumped
Programmable parameters Inlet port, outlet port, duration, temperature and an alarm for 9 programs of 20 steps
Development temp. range Ambient to 50 C
Dimensions (WxDxH) 30 x 30 x 14 cm
Weight 4.8 kg