6 + 2 Semi-Automatic Dissolution Tester
The new Mono-ShaftÔ design means you only change the stirrer inserts. The shafts are simply placed into the drive system, calibrated once and then remain there with no need for further adjustment.
Active Questions & AnswersAsk a Question
There are no current Discussions
Dissolution Testing Service ProvidersView All (16)
Documents & ManualsView All Documents
Features of PTWS 610
- Fully USP and EP compliant * 8 stirred positons * Electronic lift drive * Manual or automated tablet magazine for synchronous tablet drop available * Staggered start capability * Options – Manual or automated tablet magazine, EPE automated sampling system, ITM idividual temperature monitoring system * Interfaces – RS232 and I/O port for remote control in automated applications
General Specifications
There are no General Specifications available.
Additional Specifications
Display:
LCD Digital Display (10 x 10 cm illuminated) for RPM, temperature, time, timer and pH (optional) functions
Keyboard:
Function and alpha-numeric keys
Acoustic Signal:
Programmable acoustic signal for operator information
Interface:
1 RS-232 port
1 TTL relay port to connect a PTFC 2 Fraction collector and a pump
1 pH-probe port
Printer:
Built-in Thermo Printer, prints test log as well as OQ information
Speed control:
Adjustable from 25 rpm - 250 rpm
Accuracy:
± 2% of set speed typically < 1%
Temperature control:
750 W heater and pump system, protected against overheating and "no water" operation”, adjustable from about 25°C - 45°C, water diffuser for even water distribution all over
the bath
Accuracy:
± 0.2° C inside the water bath
Water circulation:
Water circulated through special diffusion system
pH measurement :
0.05 - 9.00 (Option)
Accuracy:
± 0.02 pH units
Number of stirred vessels:
6 vessels for buffer / medium
Additional stirred vessels:
2 vessels to take Reference Standard or refilling media
Heat-Up: Energy saving, programmable, "auto start" heater function
Sleeping Mode:
Programmable Heater Off time
Calibration:
Built-in calibration procedures for speed, temperature control and pH-probe, OQ/PQ sequence programmable including alarm indicator
Stirrer wobble:
Better than 0.2 mm total run out
System tools:
Mono-shaft stirrer design, USP Apparatus 1, 2, 6 tool adapter, cream cell, trans-dermal patch tools - each tool individually coded
Vibration:
Water bath and pump system on vibration-free mounts.
Vibration inside vessel:
vibration at the vessel flange < 0.2 mm/s @ 50 Hz
Vessel Centring:
Auto centring inside the bath cover, easily aligned bath using centring tools
Test Vessels:
1 litre USP vessels supplied with individual number coding