PHCBi - CytoGrow COMPACT Incubators

Manufactured by  PHCBi
Follow this Equipment

Panasonic CytoGROW Compact Series CO2 incubators offer precise CO2 and O2 control in a compact, space saving design...

Panasonic CytoGROW Compact Series CO2 incubators offer precise CO2 and O2 control in a compact, space saving design allowing up to three models to be stacked, making them ideal for in vitro simulation of an in vivo environment in a laboratory environment. When arrayed in triple stackable configurations, this unit meets GLP requirements by helping to avoid patient-to-patient cross contamination in individualized patient-specific clinical incubation applications.
Active Questions & AnswersAsk a Question

There are no current Discussions

CO2 Incubator Service ProvidersView All (28)

Request Support
Documents & Manuals

There are no Documents or Manuals available.

Features of CytoGrow COMPACT Incubators

Main Features:

  • Compact and Stackable
  • Active Background Control with InCu-saFe
  • SafeCell UV Protection (Optional)
  • Direct Heat and Air Jacket System
  • Integrated Microprocessor Control

Preventive Contamination Control

The unique InCu-saFe® copper-enriched stainless steel chamber with coved corners provides continuous background contamination control, inhibiting bacteria and other contaminant growth on the surface.

Ergonomic Design

The Panasonic CytoGROW Compact series incubators provide optimal configurability with its low-profile and ability to be triple-stacked. With user-friendly access built in to the door-mounted control panel and a field reversible door that can be set up for right or left-swing door openings, use your laboratory space more efficiently than ever.

Improved Temperature Stability with D.H.A. System

P.I.D. controlled 3-way heaters plus Panasonic’s proprietary D.H.A. (Direct Heat and Air jacket) provides a high-precision temperature environment, and minimizes the risk of condensation at all times.

Water Level Sensor

The humidity pan has an optical water level sensor to warn of low water level.

Automatic Gas Cylinder Switchover System

This system automatically switches from the primary to secondary gas cylinder when the O2 gas level does not change while an injection valve is open.

Contamination Control

InCu-saFe® copper-enriched stainless steel chamber construction helps inhibit the growth of mycoplasma, bacteria, molds, yeasts and fungi without costly HEPA air scrubbers, which can accumulate contaminants in the chamber. Disruptive, high temperature decontamination schemes that can actually encourage growth of heat resistant thermophilic and hyperthermophilic microorganisms in vitro are also avoided.

High Performance In Vitro Modeling

Stable temperature, humidity and CO2 density are achieved through a combination of performance systems supervised by a centralized microprocessor controller complete with alarm, programming, calibration and diagnostic protocols exportable to a remote database.

General Specifications
Depth 22.3 in
Height56.8 in
Width 18.9 in
Chamber Volume1.7 cu. ft.
Weight108 lbs
Maximum Shelf Load8.8 lbs
Shelf Dimensions6 Max - 12.2 “x 12.2 “x 0.5” / 310 x 310 x 12 mm
Additional Specifications

Decontamination:  
inCu-saFe®/ Optional SafeCell UV  
Interior Dimensions:  
13.8 X 14.9 X 14.8 (350 X 378 X 375 mm) 
Temperature Control Range:  
+5ºC above ambient to 50°C (Ambient temperature 50C to 350C 
Temperature Control Uniformity Deviation:  
±0.25°C (in 25°C ambient, setting 37°C, 5% CO2, no load 
Co2 Control:  
Thermal Conductivity  
Co2 Control Range And Deviation:  
0 to 20% ±0.15% (in 25°C ambient, setting 37°C, 5% CO2, no load) 
O2 Control And Range:  
Zirconia sensor with P.I.D.Recovery  
Interior Humidity:  
95% RH at 37°C through evaporation via DHA heating system, optional reflective/deflective optical low-water sensor 
Access Port:  
Single, 1.18/30 mm with silicone (non-VOC) stoppers (upper-left rear)

Looking for New or Used Equipment?

Shop on LabX

Shop on Labx.com