Rigaku - AutoMATE II
Micro-area X-ray residual stress measurement system. Highly accurate micro area residual stress with both iso- and side-inclination methods
Residual stress may be created during the manufacturing process of a material, or it may accumulate in a structure over many years in operation. In either case, this stress can have a serious negative effect on a product's quality, durability and lifetime. Accurate detection of residual stress is an important element of the quality control process and helps predict the service lifetime of products.
In the past, if you wanted to make highly accurate residual stress measurements, you had to use an R&D diffractometer because of the accuracy of the goniometer. However this restricts the weight and size of the samples you can measure. On the other hand, dedicated laboratory and factory-floor residual stress analyzers suffer from reduced accuracy due to the nature of their mechanical designs, while, in their favor, they have the flexibility of measuring large and heavy parts.
With the AutoMATE II, you now have the best of both worlds. Large and heavy parts (30 kg with standard manual Z stage; 20 kg with optional automated XYZ stage) can be measured with high accuracy. This is possible because the X-ray source and detector arm are mounted on a highly accurate two-axis goniometer that can position them relative to the measurement site and perform scans with an accuracy of 0.1 microns when using the automated XYZ stage.
The most advanced new feature of the AutoMATE II lies in an innovative new X-ray detector. The detector used in the AutoMATE II is the D/teX Ultra1000, an electronic Si strip detector that has high dynamic range, high sensitivity, and good energy resolution, as well as not requiring any consumable gas.
Active Questions & AnswersAsk a Question
There are no current Discussions
Elemental Analyzers Service ProvidersView All (6)
Documents & Manuals
There are no Documents or Manuals available.
Features of AutoMATE II
- Highly accurate goniometer allows for true micro-area residual stress measurement.
- Automatic mapping measurements with teaching function.
- Large and heavy samples are measured with high accuracy.
- An X-ray radiation enclosure with interlock system automatically locks the enclosure door when the X-ray shutter is open.
- The measurement position is adjusted by a CCD camera equipped with a microscope having a zoom function.
- The two-axis goniometer system allows for both iso-inclination ?and side-inclination methods automatically without readjustment of the sample position.
General Specifications
There are no General Specifications available.
Additional Specifications
[u]X-ray generator[/u]
Maximum power:
3 kW (Tube voltage 20 - 50 kV, Tube current 2 - 50 mA)
Stability:
±0.03 % (Power fluctuation within ±10%)
X-ray tube:
Standard: Cr (Maximum load 2 kW), Effective focus size 1 × 10 mm2 (N.F.), Short type
Option: Cu (2 kW), Co (1.8 kW), Fe (1.5 kW), V (0.3 kW)
[u]Goniometer[/u]
2Θ scanning range:
2Θ = 98° to 168° (Central angle range of D/teX Ultra 1000 2Θc = 108° to 158°)
Ψ angle range:
Ψ = 0° to +60° (at maximum)
Oscillation range:
Ψp = ±1° to ±10°
Incident collimator:
Standard: Φ150 Μm, Φ1 mm
Option: Φ30 Μm, Φ50 Μm, Φ100 Μm, Φ300 Μm, Φ500 Μm, Φ2 mm, Φ4 mm
Distance:
X-ray source - sample: 265 mm
Sample - detector: 210 mm
[u]Sample stage[/u]
Standard: Manual Z stage
Lab. jack (Model LJA-16223)
Maximum sample space: 720 mm (W) × 560 mm (D) × 540 mm (H)
Stage dimensions: 160 mm × 220 mm
Maximum load: 30 kg
Option: Auto XYZ stage
[i]Maximum sample space:[/i]
720 mm (W) × 560 mm (D) × 335 mm (H)
[i]Stroke:[/i]
X-Y axis = ±50 mm, Z axis = -5 mm to + 35 mm
[i]Stage dimensions:[/i]
150 mm × 150 mm
[i]Maximum load:[/i]
20 kg
X-ray shutter:
Rotary shutter
[u]Sample alignment system[/u]
CCD camera:
[i]Magnification:[/i]
×22 to ×135 (Field of vision 6 mm to 1 mm)
[i]Focal distance:[/i]
90 mm
[u]Detector (D/teX Ultra 1000)[/u]
Dimension:
One dimension (Semiconductor system)
Number of channels:
1024 ch
Maximum counting rate:
1 × 106 cps/ch × 1024 ch (Total 1 Gcps/all)
2Θ angle resolution:
0.02° (Strip width 75 Μm/line)
Window area:
76.8 mm × 10 mm
Size, Weight:
135 mm (W) × 95 mm (D) × 100 mm (H), 1.4 kg
KΒ filter:
Standard: V (Cr)Option: Ni (Cu), Fe (Co), Mn (Fe), Ti (V)