Molecular Devices - Axon™ Digidata® 1550A plus HumSilencer™

Manufactured by  Molecular Devices
Follow this Equipment

A low-noise data acquisition system with single-click 50/60 Hz line-frequency noise elimination in less than one second.

The digitizer that enables acquisition of patch-clamp data 
The Axon™ Digidata® 1550B plus HumSilencer® is one of the most advanced analog-to-digital signal converters on the market, offering the innovative line synchronous noise-cancelling feature, HumSilencer. This advanced feature learns, adapts, and eliminates local 50/60 Hz line-frequency noise and high-frequency harmonics from incoming signals. The digitizer works with pCLAMPTM 11 Software for data acquisition and analysis. The software supports eight channels of analog acquisition and four acquisition modes. 
 
50/60 hz noise cancellation 
HumSilencer is an advanced feature that learns, adapts, and eliminates local 50/60 Hz line-frequency noise patterns and associated high-frequency harmonics from incoming signal up to 20 V peak-to-peak. 
 
Noise canceling 
HumSilencer is not a filter and has no effect on acquired signals. It causes no signal distortion such as frequency change, amplitude attenuation, phase shift, or DC voltage change. 
 
Wide range of input signals 
Digidata 1550B digitizes a wide range of input signals from -10 to +10 V. 
 
HumSilencer adaptation 
HumSilencer automatically recognizes changes in noise amplitudes and adapts by removing noise in less than one second. 
 
Optimized for pCLAMP 11 Software 
The pCLAMP Software suite is the most widely-used patch-clamp electrophysiology data acquisition and analysis program for control and recording of voltage-clamp, current-clamp, and patch-clamp experiments. 
 
Three configurations 
Three configurations are available: Digidata 1550B4 with four HumSilencer channels, Digidata 1550B1 with one HumSilencer channel, and Digidata 1550B0 without a HumSilencer channel. 
 
Other features and benefits include: 
• Eliminate line-frequency noise 
• Record high resolution traces 
Digitize cellular network studies
Active Questions & AnswersAsk a Question

There are no current Discussions

Data Systems Service ProvidersView All (3)

Request Support
Documents & Manuals

There are no Documents or Manuals available.

Features of Axon™ Digidata® 1550A plus HumSilencer™
  • Eliminates 50/60 Hz line-frequency noise and associated high frequency harmonics in less than one second and with a single click     
  • Digitizes a wide range of input signals from -10 to +10 V     
  • Eliminates a maximum noise amplitude of up to 20 V, peak-to-peak     
  • 8 analog input channels to digitize acquired signals up to 500 kHz independently     
  • 8 analog output channels to send command voltage output independently     
  • 8 digital out channels to control the periphery equipment used in sophisticated experiments     
  • Independent analog-to-digital converters for each input channel ensure low crosstalk levels and high data acquisition rates.     
  • USB 2.0 connection allows operation on virtually any current PC computer running Microsoft Windows 7 (32-bit or 64-bit) Operating System, including laptops.     
  • 16-bit resolution data     
  • State-of-the-art signal-to-noise ratio     
  • All signal connectors on the front panel for ease of access and maintenance of the electrophysiology set-up     
  • Rack-mount case for tidy installation of all components in the electrophysiology set-up     
  • User guide written by scientific consultants, with the assistance of Axon Instruments staff, for straightforward start-up and as an in-depth reference.
General Specifications

There are no General Specifications available.

Additional Specifications

HumSilencer     
Maximum input signal (total of noise + signal):      
±10 V     
Maximum noise amplitude:      
20 V, peak-to-peak (on a 0 V signal)     
Noise cancellation:      
Line-frequency (50 Hz / 60 Hz) and harmonics to 10 kHz     
Cancellation response time:      
1 MΩ     
Gain value:      
1     
     
Analog Outputs     
Number of channels:      
8     
Number of DACs:      
8     
Sampling rates:      
1 Hz - 500 kHz     
Resolution:      
16-bit     
Output range:      
-10.000 V to +10.000 V     
Output impedance:      
< 0.5 Ω     
Output short circuit to signal ground:      
±25 mA     
     
Digital Outputs     
Number of bits:      
8 (of 16) supported in software     
SCOPE:      
Dedicated trigger output     
Output driver:      
Advanced CMOS (AC) compatible     
Output current:      
±4 mA source, cable     
Type:      
USB 2.0 braided     
Length:      
3 meters     
     
Trigger Inputs     
Input type:      
TTL compatible     
TAG:      
Rising-edge sensitive     
START:      
Rising-edge sensitive